Literature DB >> 11720348

On the accuracy of monomer/polymer gel dosimetry in the proximity of a high-dose-rate 192Ir source.

Y De Deene1, N Reynaert, C De Wagter.   

Abstract

The aim of this work was to investigate the applicability of MR-based polymer gel dosimetry to measure the absorbed dose distribution at short distance from an iridium-192 brachytherapy point source. In this paper, different methodological problems that may result in significant errors in the measured dose distribution are discussed. First of all the extent to which physicochemical mechanisms alter the dose response is discussed. The most important among these are the oxygen permeability of the catheter material and monomer-diffusion-related effects during irradiation. The effect of oxygen on the dose-R2 curve has been determined quantitatively and an oxygen map is performed using a well-defined external irradiation beam. The effect of diffusion of monomer during irradiation has been simulated. Another contribution of errors is related to magnetic susceptibility differences between the catheter and the gel during scanning the irradiated gel. The magnetic field distortion has been mapped by using both an experimental MRI technique and by simulation. Moreover, in constructing a dose-versus-distance curve by polar averaging, the sensitivity to the exact distance between source and point of measurement and to partial volume effects is illustrated. An optimization program is proposed to determine the location of the source on a sub-pixel scale.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11720348     DOI: 10.1088/0031-9155/46/11/304

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  7 in total

Review 1.  Air kerma and absorbed dose standards for reference dosimetry in brachytherapy.

Authors:  T Sander
Journal:  Br J Radiol       Date:  2014-07-07       Impact factor: 3.039

Review 2.  Polymer gel dosimetry.

Authors:  C Baldock; Y De Deene; S Doran; G Ibbott; A Jirasek; M Lepage; K B McAuley; M Oldham; L J Schreiner
Journal:  Phys Med Biol       Date:  2010-02-11       Impact factor: 3.609

3.  A novel thin NIPAM gel cassette dosimeter for photon-beam radiotherapy.

Authors:  Ling-Ling Hsieh; Kai-Yuan Cheng; Bor-Tsung Hsieh
Journal:  PLoS One       Date:  2012-03-12       Impact factor: 3.240

Review 4.  Three-dimensional radiation dosimetry using polymer gel and solid radiochromic polymer: From basics to clinical applications.

Authors:  Yoichi Watanabe; Leighton Warmington; N Gopishankar
Journal:  World J Radiol       Date:  2017-03-28

5.  Basic Properties of a New Polymer Gel for 3D-Dosimetry at High Dose-Rates Typical for FFF Irradiation Based on Dithiothreitol and Methacrylic Acid (MAGADIT): Sensitivity, Range, Reproducibility, Accuracy, Dose Rate Effect and Impact of Oxygen Scavenger.

Authors:  Muzafar Khan; Gerd Heilemann; Wolfgang Lechner; Dietmar Georg; Andreas Georg Berg
Journal:  Polymers (Basel)       Date:  2019-10-19       Impact factor: 4.329

Review 6.  Radiation Dosimetry by Use of Radiosensitive Hydrogels and Polymers: Mechanisms, State-of-the-Art and Perspective from 3D to 4D.

Authors:  Yves De Deene
Journal:  Gels       Date:  2022-09-19

7.  Physics-aspects of dose accuracy in high dose rate (HDR) brachytherapy: source dosimetry, treatment planning, equipment performance and in vivo verification techniques.

Authors:  Antony Palmer; David Bradley; Andrew Nisbet
Journal:  J Contemp Brachytherapy       Date:  2012-06-30