Transvenous biventricular defibrillation halves energy requirements in patients.

C Butter1, E Meisel, J Tebbenjohanns, L Engelmann, E Fleck, B Schubert, S Hahn, D Pfeiffer.   

Abstract

BACKGROUND: Defibrillation thresholds (DFT) with standard implantable cardioverter-defibrillator leads in the right ventricle (RV) may be determined by weak shock field intensity in the myocardium of the left ventricle (LV). Adding a shocking electrode in a coronary vein on the middle of the LV free wall, thereby establishing biventricular defibrillation, substantially reduced defibrillation requirements in animals. We investigated the feasibility of this approach in 24 patients receiving an implantable cardioverter-defibrillator using a prototype over-the-wire temporary LV defibrillation lead. METHODS AND
RESULTS: The LV lead was inserted through the coronary sinus, using a guide catheter and guidewire, into a posterior or lateral coronary vein whose location was determined by retrograde venography. Paired DFT testing compared a standard system (RV to superior vena cava plus can emulator [SVC+Can], 60% tilt biphasic shock) to a system including the LV lead. The biventricular system was tested with a dual-shock waveform (20% tilt monophasic shock from LV-->SVC+Can, then 60% tilt biphasic shock from RV-->SVC+Can). Twenty patients completed DFT testing. Venography and LV lead insertion time was 46+/-40 minutes. The biventricular system reduced mean DFT by 45% (8.9+/-1.1 J versus 4.9+/-0.5 J, P<0.001). Twelve patients (60%) had a standard system DFT >/=8 J, and the biventricular system gave a lower DFT in all patients. There were no adverse events related to the use of the LV lead, which was removed after testing.
CONCLUSIONS: Internal defibrillation using a transvenously inserted LV lead is feasible, produces significantly lower DFTs, and seems safe under the conditions tested. Biventricular defibrillation may be a useful option for reducing DFTs or could be added to an LV pacing lead for heart failure.

Entities:  

Mesh:

Year:  2001        PMID: 11714646     DOI: 10.1161/hc4601.099463

Source DB:  PubMed          Journal:  Circulation        ISSN: 0009-7322            Impact factor:   29.690


  6 in total

Review 1.  New technologies of internal defibrillation.

Authors:  Derek J Dosdall; Raymond E Ideker
Journal:  J Interv Card Electrophysiol       Date:  2005-08       Impact factor: 1.900

2.  Additional coronary sinus defibrillation lead with a right pectoral ICD and high DFT : a case report.

Authors:  M Strauss; T Kleemann; U Weisse; F-U Sack; R Zahn
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2011-04-10

Review 3.  Cardiac resynchronization: a brief synopsis part I: patient selection and results from clinical trials.

Authors:  David M Kalinchak; Mark H Schoenfeld
Journal:  J Interv Card Electrophysiol       Date:  2003-10       Impact factor: 1.900

4.  The use of telescoping guide catheters for coronary sinus cannulation and sub-selecting tributaries in left ventricular lead placement.

Authors:  Jürgen Vogt; Torsten Schwarz; Daniel Gras; Johannes Sperzel; Philippe Ritter; Willem de Voogt; Jean-Pierre Cebron; Martin Seifert; Bruce Tockman; Bernd Schubert; Eric Johnson; Annette Doelger; Thierry Pochet; Elisabeth Mouton; Christian Butter
Journal:  J Interv Card Electrophysiol       Date:  2007-07-07       Impact factor: 1.900

5.  Comparative prospective randomized efficacy testing of different guiding catheters for coronary sinus cannulation in heart failure patients.

Authors:  Christian Butter; Daniel Gras; Philippe Ritter; Christoph Stellbrink; Eckart Fleck; Bruce Tockman; Bernd Schubert; Thierry Pochet; Willem deVoogt
Journal:  J Interv Card Electrophysiol       Date:  2003-12       Impact factor: 1.900

6.  Extended charge banking model of dual path shocks for implantable cardioverter defibrillators.

Authors:  Derek J Dosdall; James D Sweeney
Journal:  Biomed Eng Online       Date:  2008-08-01       Impact factor: 2.819

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.