Relation between corneal thickness and intraocular pressure measurement by noncontact and applanation tonometry.

O F Recep1, H Hasiripi, N Cağil, H Sarikatipoğlu.   

Abstract

PURPOSE: To assess whether there is a significant difference in intraocular pressure (IOP) measurements between noncontact and applanation methods and if so, whether the difference is correlated with corneal thickness.
SETTING: Işik Eye Clinic, Ankara, Turkey.
METHODS: This prospective study comprised 120 eyes of 60 randomly selected patients. Noncontact tonometry, applanation tonometry, and pachymetry were performed in each eye. The results were compared and an analysis was performed to determine whether there was a correlation between the differences in IOP measurements and corneal thickness in the entire group and in subgroups composed of the halves and quarters of the group.
RESULTS: The mean IOP in the entire group was 20.38 mm Hg +/- 4.97 (SD) with the noncontact tonometer and 18.84 +/- 4.47 mm Hg with the Goldmann applanation tonometer. The difference between the measurements with the noncontact method and those with the applanation method was statistically significant (P <.01) except in cases having a corneal thickness between 513 microm and 539 microm (P >.01). There was a positive correlation between corneal thickness and the difference in measurements by noncontact and applanation tonometry, and the magnitude of correlation was greater in cases with thicker corneas.
CONCLUSION: The reliability of tonometers decreased with increasing corneal thickness, in which case higher readings were found, especially with the noncontact tonometer.

Entities:  

Mesh:

Year:  2001        PMID: 11709252     DOI: 10.1016/s0886-3350(01)00900-2

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  10 in total

1.  The influence of central corneal thickness and age on intraocular pressure measured by pneumotonometry, non-contact tonometry, the Tono-Pen XL, and Goldmann applanation tonometry.

Authors:  P-A Tonnu; T Ho; T Newson; A El Sheikh; K Sharma; E White; C Bunce; D Garway-Heath
Journal:  Br J Ophthalmol       Date:  2005-07       Impact factor: 4.638

Review 2.  The effect of caffeine on intraocular pressure: a systematic review and meta-analysis.

Authors:  Mao Li; Min Wang; Wenyi Guo; Jiajian Wang; Xinghuai Sun
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2010-08-13       Impact factor: 3.117

3.  Clinical investigation of the effect of topical anesthesia on intraocular pressure.

Authors:  Turki M Almubrad; Kelechi C Ogbuehi
Journal:  Clin Ophthalmol       Date:  2007-09

4.  [Methods of measuring intraocular pressure independently of central corneal thickness].

Authors:  A Hager; W Wiegand
Journal:  Ophthalmologe       Date:  2008-09       Impact factor: 1.059

5.  Central corneal thickness and intraocular pressure in Malay children.

Authors:  Fatemeh Heidary; Reza Gharebaghi; Wan Hazabbah Wan Hitam; Nyi Nyi Naing; Nadiah Wan-Arfah; Ismail Shatriah
Journal:  PLoS One       Date:  2011-10-05       Impact factor: 3.240

6.  Comparison of intraocular pressure adjusted by central corneal thickness or corneal biomechanical properties as measured in glaucomatous eyes using noncontact tonometers and the Goldmann applanation tonometer.

Authors:  Kiyoshi Yaoeda; Atsushi Fukushima; Motohiro Shirakashi; Takeo Fukuchi
Journal:  Clin Ophthalmol       Date:  2016-05-11

7.  Distribution and associations of intraocular pressure in 7- and 12-year-old Chinese children: The Anyang Childhood Eye Study.

Authors:  Shuning Li; Shi-Ming Li; Xiao-Lei Wang; Meng-Tian Kang; Luo-Ru Liu; He Li; Shi-Fei Wei; An-Ran Ran; Siyan Zhan; Ravi Thomas; Ningli Wang
Journal:  PLoS One       Date:  2017-08-17       Impact factor: 3.240

8.  Comparison of intraocular pressure as measured by three different non-contact tonometers and goldmann applanation tonometer for non-glaucomatous subjects.

Authors:  Seung Pil Bang; Chong Eun Lee; Yu Cheol Kim
Journal:  BMC Ophthalmol       Date:  2017-11-02       Impact factor: 2.209

9.  Comparison of variation of intraocular pressure in noncontact tonometry in patients subjected to phacoemulsification and trabeculectomy with phacoemulsification.

Authors:  Francisco Wellington Rodrigues; Henrique Ferreira Pucci; Lucas Oliveira Cintra; Rodrigo Egídio da Silva
Journal:  Clin Ophthalmol       Date:  2018-10-23

10.  Comparison of Anterior Segment Measurements with a New Multifunctional Unit and Five Other Devices.

Authors:  Bo Yi Kim; Ikhyun Jun
Journal:  Korean J Ophthalmol       Date:  2022-06-15
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.