Noninvasive cardiac output by partial CO2 rebreathing after severe chest trauma.

R A Maxwell1, J B Gibson, J B Slade, T C Fabian, K G Proctor.   

Abstract

BACKGROUND: In multiple trauma patients, early continuous cardiac output (CCO) monitoring is frequently desired but is difficult to routinely employ in most emergency departments because it requires invasive procedures. Recently, a noninvasive cardiac output (NICO) technique based on the Fick principle and partial CO2 rebreathing has shown promise under a variety of conditions. Since this method has not been tested after lung damage, we evaluated its utility in a clinically relevant model.
METHODS: Anesthetized, ventilated swine (n = 11, 35-45 kg) received a unilateral blunt trauma via a captive bolt gun followed by a 25% hemorrhage. After 60 min of shock, crystalloid resuscitation was given as needed to maintain heart rate < 100 beats/min and mean arterial pressure > 70 mm Hg. Standard CCO by thermodilution (Baxter Vigilance, Irvine, CA) was compared with NICO (Novametrix Medical Systems Inc., Wallingford, CT) for 8 hr.
RESULTS: The severity of the injury is reflected by seven deaths (average survival time = 4.25 hr). Trauma increased dead space ventilation (19%), airway resistance (30%), and lactate (3.2 mmol/L), and decreased dynamic compliance (48%) and Pao2/Fio2 (54%). In these extreme conditions, the time course and magnitude of change of CCO and NICO were superimposed. Bland-Altman analysis reveal a bias and precision of 0.01 +/- 0.69 liters/min. The linear relationship between individual CCO and NICO values was significant (p < 0.0001) and was described by the equation NICO = (0.74 +/- 0.1)CCO + (0.65 +/- 0.16 liters/min) but the correlation coefficient (r2 = 0.541) was relatively low. The cause for the low correlation could not be attributed to increased pulmonary shunt, venous desaturation, anemia, hypercapnia, increased dead space ventilation, or hyperlactacidemia.
CONCLUSION: NICO correlated with thermodilution CCO, but underestimated this standard by 26% in extreme laboratory conditions of trauma-induced cardiopulmonary dysfunction; 95% of the NICO values fall within 1.38 liters/min of CCO; and with further improvements, NICO may be useful in multiple trauma patients requiring emergency intubation during initial assessment and workup.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11706330     DOI: 10.1097/00005373-200111000-00005

Source DB:  PubMed          Journal:  J Trauma        ISSN: 0022-5282


  3 in total

1.  Tracking changes in cardiac output: methodological considerations for the validation of monitoring devices.

Authors:  Pierre Squara; Maurizio Cecconi; Andrew Rhodes; Mervyn Singer; Jean-Daniel Chiche
Journal:  Intensive Care Med       Date:  2009-07-11       Impact factor: 17.440

2.  Effects of reduced rebreathing time, in spontaneously breathing patients, on respiratory effort and accuracy in cardiac output measurement when using a partial carbon dioxide rebreathing technique: a prospective observational study.

Authors:  Kazuya Tachibana; Hideaki Imanaka; Muneyuki Takeuchi; Tomoyo Nishida; Yuji Takauchi; Masaji Nishimura
Journal:  Crit Care       Date:  2005-09-07       Impact factor: 9.097

Review 3.  Equipment review: new techniques for cardiac output measurement--oesophageal Doppler, Fick principle using carbon dioxide, and pulse contour analysis.

Authors:  Christine Berton; Bernard Cholley
Journal:  Crit Care       Date:  2002-04-25       Impact factor: 9.097

  3 in total