Literature DB >> 11696904

3D volume-ablation rate and thermal side effects with the Er:YAG and Nd:YAG laser.

A Mehl1, L Kremers, K Salzmann, R Hickel.   

Abstract

OBJECTIVES: The aim of this investigation was to determine the influence of a variety of parameters on the effectiveness of hard substance ablation and the thermal side effects when using Er:YAG laser (Key I and II, KaVo) and Nd:YAG laser (SunLase 800, Sunrise Technologies/Orbis).
METHODS: For this study, ablation and temperature measurements were carried out on 170 dentin slices and 170 extracted teeth via computer-controlled cavity preparation. The Er:YAG laser settings varied from 250-400 mJ/pulse, 3-15 pps and 20-180 s processing time, and in the case of the Nd:YAG laser from 83-100 mJ/pulse, 10-20 pps, and 20-260 s processing time. The ablation rate was measured volumetrically via a 3D sensor. Temperatures were measured for each setting both on the dentin slice and in the pulp of the extracted teeth. The results were analyzed using a t-test for independent samples and a one-way ANOVA (Bonferroni). Also a liner regression analysis was done using Pearson's coefficient.
RESULTS: The results show that with the Er:YAG laser, in combination with water-spray cooling, an effective 3D ablation rate (up to 0.017 mm3/pulse = 50 microns linear) can be achieved without raising the temperature of the surrounding tissue. In the case of the Nd:YAG laser, no measurable ablation rate was evident without conditioning of the dentin surface and, in the case of conditioning with black ink, a low ablation rate (0.00004 mm3/pulse = linear 0.2 micron/pulse) was found. SIGNIFICANCE: In contrast to the Er:YAG laser, it is apparent, that with the Nd:YAG laser from a total energy of 80 J onwards, the rise in temperature in the pulp is above 8 degrees C. For that reason, the use of the Nd:YAG laser at higher total energies is not recommended. The temperature rise with the Nd:YAG laser is dependent on the direction of the dentin tubuli. Dentin tubuli running parallel to the surface prevent significant heat penetration, whereas those running in a transverse direction to the surface (= parallel to the laser beam) support the penetration of heat. This finding supports the light-propagating theory for spreading effects of laser beams in dentin.

Entities:  

Mesh:

Substances:

Year:  1997        PMID: 11696904     DOI: 10.1016/S0109-5641(97)80036-X

Source DB:  PubMed          Journal:  Dent Mater        ISSN: 0109-5641            Impact factor:   5.304


  21 in total

1.  The effect of Nd:YAG and Er,Cr:YSGG lasers on the microhardness of human dentin.

Authors:  Wael M Al-Omari; Joseph E Palamara
Journal:  Lasers Med Sci       Date:  2012-04-24       Impact factor: 3.161

2.  Effects of different application durations of ER:YAG laser on intrapulpal temperature change during debonding.

Authors:  Didem Nalbantgil; M Oguz Oztoprak; Murat Tozlu; Tülin Arun
Journal:  Lasers Med Sci       Date:  2010-06-10       Impact factor: 3.161

3.  Evolution of the role of phototherapy during endodontic decontamination.

Authors:  Omid Heidar Muhammad; Jean-Paul Rocca; Carlo Fornaini; Etienne Medioni
Journal:  Laser Ther       Date:  2015-12-30

4.  Composite filling removal with erbium:yttrium-aluminum-garnet laser: morphological analyses.

Authors:  Alessandra M Correa-Afonso; Regina G Palma-Dibb; Jesus Djalma Pécora
Journal:  Lasers Med Sci       Date:  2008-07-04       Impact factor: 3.161

5.  Precise ablation of dental hard tissues with ultra-short pulsed lasers. Preliminary exploratory investigation on adequate laser parameters.

Authors:  Marina Stella Bello-Silva; Martin Wehner; Carlos de Paula Eduardo; Friedrich Lampert; Reinhart Poprawe; Martin Hermans; Marcella Esteves-Oliveira
Journal:  Lasers Med Sci       Date:  2012-05-08       Impact factor: 3.161

6.  In vitro investigation of intra-canal dentine-laser beam interaction aspects: I. Evaluation of ablation capability (ablation rate and efficiency).

Authors:  Nova Hambersom Minas; Joerg Meister; Rene Franzen; Norbert Gutknecht; Friedrich Lampert
Journal:  Lasers Med Sci       Date:  2009-07-28       Impact factor: 3.161

7.  In vitro investigation of intra-canal dentine-laser beam interaction aspects: II. Evaluation of ablation zone extent and morphology.

Authors:  Nova Hambersom Minas; Norbert Gutknecht; Friedrich Lampert
Journal:  Lasers Med Sci       Date:  2009-08-29       Impact factor: 3.161

8.  Effects of different application durations of scanning laser method on debonding strength of laminate veneers.

Authors:  Mehmet Oguz Oztoprak; Murat Tozlu; Ufuk Iseri; Feyza Ulkur; Tulin Arun
Journal:  Lasers Med Sci       Date:  2011-07-12       Impact factor: 3.161

9.  Influence of water-layer thickness on Er:YAG laser ablation of enamel of bovine anterior teeth.

Authors:  Maziar Mir; Joerg Meister; Rene Franzen; Shabnam S Sabounchi; Friedrich Lampert; Norbert Gutknecht
Journal:  Lasers Med Sci       Date:  2007-10-20       Impact factor: 3.161

10.  Effect of Er:YAG laser on debonding strength of laminate veneers.

Authors:  Ufuk Iseri; Mehmet Oguz Oztoprak; Zeynep Ozkurt; Ender Kazazoglu; Tulin Arun
Journal:  Eur J Dent       Date:  2014-01
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.