BACKGROUND: Reliability cannot be achieved in a clinical laboratory through the control of accuracy in the analytical phase of the testing process alone. Indeed a "mistake" can be defined as any defect occuring during the testing process. In the analysis of clinical specimens, there are many possible preanalytical sources of error. Therefore, the application of quality system to laboratory testing requires total quality management throughout the laboratory process, including the preanalytical and postanalytical phases. ISO 9002:1994 is a model for quality assurance in production, installation, and servicing, which includes a number of clauses providing guidance for implementation in clinical laboratories. Our laboratory at King Chulalongkorn Memorial Hospital, the largest Thai Red Cross Society hospital, is the first clinical laboratory in Thailand with ISO 9002:1994 certified for the whole unit.Entities:
Year: 2001 PMID: 11696253 PMCID: PMC59663 DOI: 10.1186/1472-6890-1-5
Source DB: PubMed Journal: BMC Clin Pathol ISSN: 1472-6890
Examples of documents according to ISO 9002:1994 in our setting.
| types | examples |
| 1. Quality system procedure | Statistical technique |
| Process control | |
| Management review | |
| 2. Work instruction | Fire prevention |
| Patient education | |
| 3. Form | Data record form for preanalytical mistake |
| Client satisfaction survey form |
Example of data record form for preanalytical mistake
| Date | time | Source of | Type* and detail of | Primary corrective action | signature |
| request | identified mistake | ||||
* classified as physician's order missed, patient misidentiflcation, specimen collected in insufficient quantity, inappropriate container used, inappropriate quality of specimen, specimen lost in transit and etc.
Preanalytical mistakes in laboratory testing.
| Number | Number of | Number of | |||
| Division | of samples* | requested | preanalytical | Frequency (%) | (95% Cl) |
| tests | mistakes | ||||
| In-Patient | 298,769 | 304,906 | 426 | 0.143 | 0.129–0.156 |
| Out-Patient | 441,268 | 457,013 | 622 | 0.141 | 0.130–0.152 |
| Total | 935,896 | 941,902 | 1,048 | 0.112 | 0.105–0.119 |
* The number of requested tests were greater than the number of samples because more than one test could be performed on a sample.
Types of preanalytical mistakes in laboratory testing.
| Defects detection steps | Defects found | |
| number | Frequency(%) | |
| Wrong identification of the patient | 281 | 26.81 |
| Physician's order missed | 147 | 14.03 |
| Inappropriate container used | 6 | 0.57 |
| Inappropriate quantity of specimen | 121 | 11.55 |
| Inappropriate quality of specimen | 493 | 47.04 |