Final results of the CAPAS trial.

M Izumi1, E Tsuchikane, M Funamoto, T Kobayashi, S Sumitsuji, S Otsuji, M Sakurai, T Kobayashi, N Awata.   

Abstract

BACKGROUND: The Cutting Balloon (Interventional Technologies Inc) is a new-concept balloon that incorporates 3 to 4 blades to create sharp incisions on the luminal surface of the lesion during dilation without causing severe tearing injury to the vessel wall. It may reduce restenosis and improve clinical outcome.
METHODS: Two hundred forty-eight lesions were randomly assigned to Cutting Balloon angioplasty (CBA, 120 lesions) or conventional balloon angioplasty (PTCA, 128 lesions). Inclusion criteria were type B/C lesions (American College of Cardiology/American Heart Association classification) and reference diameter <3.0 mm by visual image on angiogram. Quantitative coronary angiography was performed before and after percutaneous coronary angioplasty and at 3-month follow-up. The primary end point was restenosis, defined as >/=50% diameter stenosis at follow-up. Clinical event rates at 1 year were assessed.
RESULTS: Baseline characteristics were similar. Reference diameter was small in both groups (2.16 vs 2.18 mm, CBA vs PTCA). Preprocedural percent diameter stenosis (%DS) was similar (69.8% vs 69.6%). However, postprocedural and follow-up %DS were lower (26.2% vs 28.9%, P =.072; 40.8% vs 47.5%, P =.011) in the CBA group. Restenosis was significantly lower (25.2% vs 41.5%, P =.009) in the CBA group. At 1 year, event-free survival was achieved in 72.8% of the CBA group and in 61.0% of the PTCA group (P =.047).
CONCLUSION: These findings suggest that CBA provides superior angiographic and clinical outcomes in comparison with PTCA in small coronary arteries.

Entities:  

Mesh:

Year:  2001        PMID: 11685163     DOI: 10.1067/mhj.2001.119129

Source DB:  PubMed          Journal:  Am Heart J        ISSN: 0002-8703            Impact factor:   4.749


  2 in total

1.  Percutaneous coronary intervention: recommendations for good practice and training.

Authors:  K D Dawkins; T Gershlick; M de Belder; A Chauhan; G Venn; P Schofield; D Smith; J Watkins; H H Gray
Journal:  Heart       Date:  2005-12       Impact factor: 5.994

Review 2.  Contemporary tools and devices for coronary calcium modification.

Authors:  Heerajnarain Bulluck; Margaret McEntegart
Journal:  JRSM Cardiovasc Dis       Date:  2022-04-05
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.