Literature DB >> 11677946

Microwave coagulation versus insertion of a second stent for occluded biliary metal stent.

I Maetani1, T Ukita, H Inoue, M Sato, Y Igarashi, Y Sakai.   

Abstract

BACKGROUND/AIMS: There is no consensus regarding optimal management of self-expandable metallic stent occlusion. We investigated the efficacy of microwave coagulation therapy for recanalization as compared to second stent placement.
METHODOLOGY: Sixty patients with malignant obstruction of the common bile duct were treated with metal stent placement from January 1992 to July 1999. Of these, 13 patients subsequently developed stent occlusion due to tumor ingrowth. We compared stent patency and patient survival rates after microwave coagulation to those after insertion of a second stent. The influence of the duration of patency of the first stent on the second stent patency was also evaluated.
RESULTS: Of the 13 patients with stent occlusion, 7 were treated with microwave coagulation therapy, and 6 with insertion of a second metal stent. In all cases, occluded stents were successfully recanalized without any complications. There was no significant difference in duration of first stent patency between the two groups. The median duration of second stent patency was prolonged in microwave-treated patients (152 days vs. 104 days, P > 0.05). The median duration of patient survival after last recanalizing procedure was also prolonged in microwave-treated patients (131 days vs. 78 days, P > 0.05). Microwave energy did not induce destruction of the stent filament.
CONCLUSIONS: Microwave coagulation did not offer significantly longer duration of stent patency and patient survival compared to insertion of a second metal stent. However, this procedure is safe, feasible, and certainly as good as a second stent placement. It may be an alternative to insertion of a second stent within the occluded stent.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11677946

Source DB:  PubMed          Journal:  Hepatogastroenterology        ISSN: 0172-6390


  2 in total

Review 1.  Management of occluded metal stents in malignant biliary obstruction: similar outcomes with second metal stents compared to plastic stents.

Authors:  Tilak Shah; Svetang Desai; Mahfuzul Haque; Hassan Dakik; Deborah Fisher
Journal:  Dig Dis Sci       Date:  2012-06-26       Impact factor: 3.199

2.  Initial study of sediment antagonism and characteristics of silver nanoparticle-coated biliary stents in an experimental animal model.

Authors:  Yigeng Tian; Mingfeng Xia; Shuai Zhang; Zhen Fu; Qingbin Wen; Feng Liu; Zongzhen Xu; Tao Li; Hu Tian
Journal:  Int J Nanomedicine       Date:  2016-04-28