Literature DB >> 11672630

Validation of estimated 3D temperature maps during hepatic cryo surgery.

E Samset1, T Mala, B Edwin, I Gladhaug, O Søreide, E Fosse.   

Abstract

A simple model for estimating temperature distribution within the frozen region during cryo ablation was tested for accuracy. Freezing experiments were conducted in both ex vivo and in vivo porcine livers. Temperature was measured during freezing using a fiber-optic temperature sensor. Three-dimensional MR images were obtained at the end of each freezing cycle. From the MR image volumes, three-dimensional temperature maps were calculated numerically using a simplified bio-heat model. Estimated temperatures were compared to measured temperatures. The median difference between measured and estimated temperature was 3.03 degrees C. The median distance from a sensor element to the closest point on a isotherm surface with the corresponding estimated temperature was 0.70 mm. The accuracy of this model is acceptable. Temperature maps as outlined here may be used for monitoring of cryotherapy in order to increase clinical effectiveness.

Mesh:

Year:  2001        PMID: 11672630     DOI: 10.1016/s0730-725x(01)00389-7

Source DB:  PubMed          Journal:  Magn Reson Imaging        ISSN: 0730-725X            Impact factor:   2.546


  3 in total

Review 1.  Fiber Optic Sensors for Temperature Monitoring during Thermal Treatments: An Overview.

Authors:  Emiliano Schena; Daniele Tosi; Paola Saccomandi; Elfed Lewis; Taesung Kim
Journal:  Sensors (Basel)       Date:  2016-07-22       Impact factor: 3.576

Review 2.  Optical Fiber Temperature Sensors and Their Biomedical Applications.

Authors:  Paulo Roriz; Susana Silva; Orlando Frazão; Susana Novais
Journal:  Sensors (Basel)       Date:  2020-04-09       Impact factor: 3.576

3.  Fiber Bragg Grating Sensors for Performance Evaluation of Fast Magnetic Resonance Thermometry on Synthetic Phantom.

Authors:  Martina De Landro; Jacopo Ianniello; Maxime Yon; Alexey Wolf; Bruno Quesson; Emiliano Schena; Paola Saccomandi
Journal:  Sensors (Basel)       Date:  2020-11-12       Impact factor: 3.576

  3 in total