Assessment of Doppler ultrasound to measure systolic and diastolic blood pressures in infants and young children.

R K Whyte, A M Elseed, C B Fraser, E A Shinebourne, M de Swiet.   

Abstract

A recently developed instrument uses the Doppler shift technique to detect vessel wall movement, and it has been suggested that in conjunction with a conventional sphygmomanometer systolic and diastolic blood pressures can be measured. A controlled study was carried out in 20 children recovering from cardiac surgery where direct intra-arterial measurements (one observer) were compared with independent measurements using the Doppler instrument (2 observers). Systolic pressures measured directly and by Doppler technique correlated well and there was no significant difference between intra-arterial and indirect measurements whether the latter were taken by doctors or by nurses. In contrast, direct and indirect diastolic pressure measurements correlated poorly and were significantly overestimated with a mean difference of 6-25 mmHg (range +25 to -10) for doctors, and 4-25 mmHg (range +20 to -10) for nurses. Thus, the instrument adequately measured systolic blood pressure, but in our hands did not give precise measurements for diastolic blood pressure.

Entities:  

Mesh:

Year:  1975        PMID: 1167067      PMCID: PMC1544615          DOI: 10.1136/adc.50.7.542

Source DB:  PubMed          Journal:  Arch Dis Child        ISSN: 0003-9888            Impact factor:   3.791


  2 in total

1.  Accuracy of an ultrasound blood pressure instrument in neonates, infants, and children.

Authors:  H M Hochberg; M B Saltzman
Journal:  Curr Ther Res Clin Exp       Date:  1971-07

2.  Assessment of techniques for measurement of blood pressure in infants and children.

Authors:  A M Elseed; E A Shinebourne; M C Joseph
Journal:  Arch Dis Child       Date:  1973-12       Impact factor: 3.791

  2 in total
  7 in total

Review 1.  Blood pressure.

Authors:  M J Dillon
Journal:  Arch Dis Child       Date:  1988-04       Impact factor: 3.791

Review 2.  Clinical diagnosis by transcutaneous Doppler ultrasound.

Authors:  R K Wyse
Journal:  Postgrad Med J       Date:  1982-04       Impact factor: 2.401

3.  Elevated systolic blood pressure in preterm very-low-birth-weight infants ≤3 years of life.

Authors:  Andrea F Duncan; Roy J Heyne; Janet S Morgan; Naveed Ahmad; Charles R Rosenfeld
Journal:  Pediatr Nephrol       Date:  2011-03-13       Impact factor: 3.714

4.  Blood pressure in infants children and adolescents.

Authors:  A J Moss
Journal:  West J Med       Date:  1981-04

5.  Automatic noninvasive measurement of systolic blood pressure using photoplethysmography.

Authors:  Meir Nitzan; Amikam Patron; Zehava Glik; Abraham T Weiss
Journal:  Biomed Eng Online       Date:  2009-10-26       Impact factor: 2.819

6.  Clinical evaluation and comparison of the Infrasonde, Arteriosonde, and mercury sphygmomanometer in measurement of blood pressure in children.

Authors:  J M Savage; M J Dillon; J F Taylor
Journal:  Arch Dis Child       Date:  1979-03       Impact factor: 3.791

7.  Comparison of systolic blood pressure values obtained by photoplethysmography and by Korotkoff sounds.

Authors:  Meir Nitzan; Yair Adar; Ellie Hoffman; Eran Shalom; Shlomo Engelberg; Iddo Z Ben-Dov; Michael Bursztyn
Journal:  Sensors (Basel)       Date:  2013-10-31       Impact factor: 3.576

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.