Effect of stimulation parameters on electrically evoked auditory brainstem responses.

G A Tavartkiladze1, L A Potalova, A V Kruglov, A Belov.   

Abstract

It is evident that the conventional technique for cochlear implant adjustment is not suitable for children in their first years of life. In order to find a solution to this problem, the possibility of electrically evoked auditory brainstem response (EABR) recording was investigated. EABRs were recorded in 9 patients with the Nucleus multichannel cochlear implant. The main problems that have to be solved during EABR recording in cochlear implantees are: i) EABR distortion due to the stimulus artefact: and ii) difference in the stimulus presentation rate during EABR registration (low pulse rate) and conventional psychophysical threshold estimation (high pulse rate) in cochlear implant patients. The influence of stimulus artefact on the recording results was minimized by setting the implant to the widest amplifier frequency band and by zeroing the initial segment containing the stimulus artefact with subsequent zero-phase digital filtering. The dependence of the EABR amplitude and latency on the stimulus intensity, width, electrode location and interstimulus interval was investigated. It was concluded that despite the difference revealed between absolute values of EABR thresholds and psychophysical threshold levels, it is possible to calculate implant adjustment parameters based on the EABR data with the proper correction applied.

Entities:  

Mesh:

Year:  2000        PMID: 11603775     DOI: 10.1080/000164800750000946

Source DB:  PubMed          Journal:  Acta Otolaryngol        ISSN: 0001-6489            Impact factor:   1.494


  3 in total

1.  Intra-Operative Neural Response Telemetry and Acoustic Reflex Assessment using an Advance-In-Stylet Technique and Modiolus-Hugging: A prospective cohort study.

Authors:  Abdulrahman Hagr
Journal:  Sultan Qaboos Univ Med J       Date:  2011-08-15

2.  The polarity sensitivity of the electrically stimulated human auditory nerve measured at the level of the brainstem.

Authors:  Jaime A Undurraga; Robert P Carlyon; Jan Wouters; Astrid van Wieringen
Journal:  J Assoc Res Otolaryngol       Date:  2013-03-12

3.  Examination and Comparison of Electrically Evoked Compound Action Potentials and Electrically Evoked Auditory Brainstem Response Results of Children with Cochlear Implantation without Inner Ear Anomaly.

Authors:  Seda Bayrak; Başak Mutlu; Günay Kırkım; Bülent Şerbetçioğlu
Journal:  Turk Arch Otorhinolaryngol       Date:  2019-06-27
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.