Experimental and finite element comparison of various fixation designs in combined loads.

A Shirazi-Adl1, O Patenaude, M Dammak, D Zukor.   

Abstract

The short- and long-term successes of tibial cementless implants depend on the initial fixation stability often provided by posts and screws. In this work, a metallic plate was fixed to a polyurethane block with either two bone screws, two smooth-surfaced posts, or two novel smooth-surfaced posts with adjustable inclinations. For this last case, inclinations of 0, 1.5, and 3 deg were considered following insertion. A load of 1031 N was eccentrically applied on the plate at an angle of approximately 14 deg, which resulted in a 1000 N axial compressive force and a 250 N shear force. The response was measured under static and repetitive loading up to 4000 cycles at 1 Hz. The measured results demonstrate subsidence under load, lift-off on the unloaded side, and horizontal translation of the plate specially at the loaded side. Fatigue loading increased the displacements, primarily during the first 100 cycles. Comparison of various fixation systems indicated that the plate with screw fixation was the stiffest with the least subsidence and liftoff. The increase in post inclination from 0 to 3 deg stiffened the plate by diminishing the liftoff. All fixation systems demonstrated deterioration under repetitive loads. In general, the finite element predictions of the experimental fixation systems were in agreement with measurements. The finite element analyses showed that porous coated posts (modeled with nonlinear interface friction with and without coupling) generated slightly less resistance to liftoff than smooth-surfaced posts. In the presence of porous coated posts, Coulomb friction greatly overestimated the rigidity by reducing the liftoff and subsidence to levels even smaller than those predicted for the design with screw fixation. The sequence of combined load application also influenced the predicted response. Finally, the finite element model incorporating measured interface friction and pull-out responses can be used for the analysis of cementless total joint replacement systems during the post-operation period.

Entities:  

Mesh:

Year:  2001        PMID: 11601722     DOI: 10.1115/1.1395572

Source DB:  PubMed          Journal:  J Biomech Eng        ISSN: 0148-0731            Impact factor:   2.097


  3 in total

1.  Fatigue characterization of a polymer foam to use as a cancellous bone analog material in the assessment of orthopaedic devices.

Authors:  V Palissery; M Taylor; M Browne
Journal:  J Mater Sci Mater Med       Date:  2004-01       Impact factor: 3.896

2.  Development of a subject-specific model to predict the forces in the knee ligaments at high flexion angles.

Authors:  Zhaochun Yang; Alexis C Wickwire; Richard E Debski
Journal:  Med Biol Eng Comput       Date:  2010-06-29       Impact factor: 2.602

3.  Finite element analysis of the tibial bone graft in cementless total knee arthroplasty.

Authors:  Koji Totoribe; Etsuo Chosa; Go Yamako; Hiroaki Hamada; Koki Ouchi; Shutaro Yamashita; Gang Deng
Journal:  J Orthop Surg Res       Date:  2018-05-16       Impact factor: 2.359

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.