| Literature DB >> 11600310 |
T Schenk1, H Irth, G Marko-Varga, L E Edholm, U R Tjaden, J van der Greef.
Abstract
An on-line liquid chromatography-immunochemical detection (LC-ICD) system for the quantification of cytokines in cell extracts has been developed using a post-column continuous-flow reaction detection system using fluorescence labelled antibodies. Cytokines eluting from the micro-HPLC column react with antibodies to form fluorescent complexes. In a second step the excess of free antibody is trapped on a cytokine bound support prior to fluorescence detection. The concentration detection limit of the flow injection-ICD system was 50 pM (20 microl injection volume) for interleukin 4 (IL-4). An absolute detection limit of 1 fmol was obtained for IL-4. Similar to ICD systems for small non-protein analytes developed earlier, reaction times were in the order of 1 minute. The immobilised cytokine affinity columns can easily be regenerated and used for months. The present ICD system for interleukins 4, 6, 8 and 10 was coupled to ion exchange-, size exclusion- and reversed phase chromatography. Important parameters (reaction times, reaction conditions) were investigated to get a better understanding of post-column ICD systems for macromolecules.Entities:
Mesh:
Substances:
Year: 2001 PMID: 11600310 DOI: 10.1016/s0731-7085(01)00464-2
Source DB: PubMed Journal: J Pharm Biomed Anal ISSN: 0731-7085 Impact factor: 3.935