A new modular testing system for biomechanical evaluation of tibial intramedullary fixation devices.

C Gaebler1, A Speitling, E L Milne, S Stanzl-Tschegg, V Vécsei, L L Latta.   

Abstract

This biomechanical study was performed to evaluate a new modular, tibial testing system developed for analysis of tibial nails and their locking screws.A new testing system, consisting of five modules, was designed to simulate a tibia. For this study one module was removed to simulate a 55-mm distal tibial defect inducing maximum loading on the distal portion of the implant and locking bolts. The tibial load offsets were 23 mm proximally and 10 mm distally medial to the centreline of the tibial shaft to simulate the location of the expected resultant load during the peak loading and inversion torque on the ankle during the gait cycle. Four solid tibial nails (STN, Stryker-Howmedica-Osteonics, Kiel, Germany) were tested to static failure and 15 nails were tested dynamically. Our results showed that the solid tibial nails fractured in the testing device in the same manner and location as they do in clinical series. Evaluation of the results showed the mean fatigue limit of the STN to be 1.4 kN for 500,000 cycles with a standard deviation (S.D.) of 0.33 kN. This biomechanical study establishes a standard technique for the biomechanical testing of tibial nails, in a clinically relevant manner, avoiding the inconsistency of cadaver bone tests. As it is a standardised test set-up this new modular testing system could serve as a standard by which small diameter tibial nails and other devices could be evaluated and compared with other systems currently in use.

Entities:  

Mesh:

Year:  2001        PMID: 11600118     DOI: 10.1016/s0020-1383(01)00044-4

Source DB:  PubMed          Journal:  Injury        ISSN: 0020-1383            Impact factor:   2.586


  3 in total

1.  Implant material and design alter construct stiffness in distal femur locking plate fixation: a pilot study.

Authors:  Ulf Schmidt; Rainer Penzkofer; Samuel Bachmaier; Peter Augat
Journal:  Clin Orthop Relat Res       Date:  2013-09       Impact factor: 4.176

2.  Biomechanical comparison of six intramedullary nails, for the treatment of ex-tra-articular, proximal tibial fractures.

Authors:  G Gkouvas; F Agathangelidis; C Nakas; C David; D Sagris; G Petsatodis
Journal:  Hippokratia       Date:  2019 Apr-Jun       Impact factor: 0.471

3.  Experimental and numerical investigation into the influence of loading conditions in biomechanical testing of locking plate fracture fixation devices.

Authors:  A MacLeod; A H R W Simpson; P Pankaj
Journal:  Bone Joint Res       Date:  2018-01       Impact factor: 5.853

  3 in total