| Literature DB >> 11591900 |
C Menor1, J A Fueyo, O Escribano, C Cara, M D Fernández-Moreno, I D Román, L G Guijarro.
Abstract
The current article describes a new assay to measure thiopurine methyltransferase (TPMT) activity from red blood cells. This method is based on the measurement of the reaction product 6-methylmercaptopurine (6-MMP) by high-performance liquid chromatography (HPLC). 6-MMP is extracted by ethyl acetate with recoveries of 85%, 80%, 80%, and 92% for 50, 250, 500, and 1,000 ng/100 microL packed red blood cells, respectively. 6-MMP was identified and measured by a Zorbax CN column installed in an HPLC system. The chromatograms were resolved using a mobile phase consisting of 40 mmol/L sodium phosphate buffer (pH 3) and methanol in a gradient from 1% to 20% of methanol. Under these conditions 6-MMP is well resolved from substrates (6-mercaptopurine and S-adenosyl-L-methionine) and endogenous peaks. When the TPMT activity from 20 patients was measured by the HPLC-linked assay and the classic radiochemical method, a linear correlation was obtained between both procedures ( y = 0.99x + 0.33; x-axis, radiochemical assay; y-axis, HPLC-linked assay; r = 0.98). In conclusion, the current report describes a new, reliable, safe, and nonradioactive method to measure TPMT activity that is shorter and simpler than the previously described ones.Entities:
Mesh:
Substances:
Year: 2001 PMID: 11591900 DOI: 10.1097/00007691-200110000-00007
Source DB: PubMed Journal: Ther Drug Monit ISSN: 0163-4356 Impact factor: 3.681