| Literature DB >> 11589453 |
T Kojima1, A Ishii, K Watanabe-Suzuki, R Kurihara, H Seno, T Kumazawa, O Suzuki, Y Katsumata.
Abstract
Four general anaesthetics, sevoflurane, isoflurane, enflurane and halothane, in human whole blood, have been found measurable with very high sensitivity by capillary gas chromatography-flame ionization detection (GC-FID) with cryogenic oven trapping upon injection of headspace (HS) vapor sample. To a 7-ml vial, containing 0.48 ml of distilled water and 20 microl of internal standard solution (5 microg), a 0.5-ml of whole blood sample spiked with or without anaesthetics, was added, and the mixture was heated at 55 degrees C for 15 min. A measure of 10 ml HS vapor was injected into the GC in the splitless mode at -40 degrees C oven temperature, which was programmed up to 250 degrees C. All four peaks were clearly separated; no impurity peaks were found among their peaks. Their extraction efficiencies were about 10%. The calibration curves showed good linearity in the range of 0.5-20 microg/ml; their detection limits were 10-100 ng/ml, which are almost comparable to those by previous reports. The coefficients of intra-day and day-to-day variations were 6.5-9.8 and 7.3-17.2%, respectively. Isoflurane or enflurane was also measured from whole blood samples in which three volunteers inhaled each compound.Entities:
Mesh:
Substances:
Year: 2001 PMID: 11589453 DOI: 10.1016/s0378-4347(01)00348-6
Source DB: PubMed Journal: J Chromatogr B Biomed Sci Appl ISSN: 1387-2273