The determination of low lead levels in the bone of lead-depleted mice by graphite furnace atomic absorption spectrometry.

I Iavicoli1, G Carelli, N Castellino, G Schlemmer.   

Abstract

Low lead levels in the femurs of mice fed with a lead-depleted diet have been determined by use of electrothermal atomic absorption spectrometry with Zeeman-effect background correction. The method is based on the use of Mg(NO3)2/Pd as matrix modifier which enables significant reduction of the spectral interferences prevalent if chemical modifiers based on NH4H2PO4 with either Ca or Mg are used for samples rich in Ca3(PO4)2 matrix. The method was developed and validated by use of the NIST standard reference material 1486 bone. Bones were decomposed in a pressurized microwave-heated system using 70% nitric acid. Forty-three mice femurs, with a mass of 74.62 +/- 12.54 mg, were dissolved in concentrated nitric acid. The lead results found in SRM 1486 (1.25 +/- 0.15 microg g(-1), n = 9) were in good agreement with the certificate (1.335 +/- 0.014 microg g(-1)). Recoveries of 200 ng lead added to the SRM before or after digestion were 99.0 +/- 1.4% and 98.5 +/- 1.6%, respectively. The lead detection limit in bone samples is 0.06 microg g(-1) dry mass. This method is, therefore, suitable for the determination of very low lead levels (0.06-0.20 microg Pb kg(-1) bone) in the femurs of mice fed a diet with lead level of < 20 microg kg(-1).

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11583095     DOI: 10.1007/s002160100917

Source DB:  PubMed          Journal:  Fresenius J Anal Chem        ISSN: 0937-0633


  2 in total

1.  Binding and leakage of barium in alginate microbeads.

Authors:  Yrr A Mørch; Meirigeng Qi; Per Ole M Gundersen; Kjetil Formo; Igor Lacik; Gudmund Skjåk-Braek; Jose Oberholzer; Berit L Strand
Journal:  J Biomed Mater Res A       Date:  2012-06-14       Impact factor: 4.396

2.  Sub-Chronic Oral Exposure to Iridium (III) Chloride Hydrate in Female Wistar Rats: Distribution and Excretion of the Metal.

Authors:  Ivo Iavicoli; Luca Fontana; Antonio Bergamaschi; Marcelo Enrique Conti; Anna Pino; Daniela Mattei; Beatrice Bocca; Alessandro Alimonti
Journal:  Dose Response       Date:  2012-01-16       Impact factor: 2.658

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.