F Y Chang1, C L Lu, C Y Chen, S D Lee, D S Tsai, S E Fu.
Abstract
The present study aimed to validate the accuracy of a homemade applied potential tomography (APT) apparatus for liquid gastric emptying (GE). Twelve electrodes were placed in a circular array around the subjects. Ten electrodes in a rotated order recorded electrical current delivered from a driving pair of electrodes. Based on tomography, averaged signals of changed resistivity were constructed to display area changes. Six beakers were respectively placed into a saline-filled Perspex tank to measure their cross-sections. True beaker cross-sections are 2.01, 15.9, 18.8, 30.19, 38.48, and 63.61 cm2, respectively, whereas APT generated cross-sections were 7.9 +/- 2.9, 16.7 +/- 3.3, 22.4 +/- 4.9, 28 +/- 4.8, 48.7 +/- 7.6, 67 +/- 6.1 cm2, respectively (r = 0.98, P < 0.001). Twenty-four healthy males ingested 500-ml test solution to assess GE using both APT and scintigraphy. Only 20 (83.3%) subjects had a successful measurement. The APT half emptying time was 15.6 +/- 4.8 min, whereas scintigraphy was 21.9 +/- 6.3 min (r = 0.67, P < 0.01). In conclusion, our APT apparatus is a simple, noninvasive, and inexpensive way to assess liquid GE. Its clinical usefulness is confirmed using both phantom and human models.Entities:
Mesh:
Year: 2001 PMID: 11575434 DOI: 10.1023/a:1010614526685
Source DB: PubMed Journal: Dig Dis Sci ISSN: 0163-2116 Impact factor: 3.199