L M Killingsworth, C E Britain, L L Woodard.
Abstract
An automated continuous-flow procedure has been developed for the rapid determination of urinary proteins of plasma origin. Antiserum to whole human plasma was used as the reagent, and the antigen-antibody reactions were quantitated by nephelometry. By adding polyethylene glycol (mol wt 6000-7500) to the reaction medium, reaction time was decreased to less than 3 min; no sample blanks were required; and samples were analyzed at a rate of 70 per hour. Recovery studies yielded an average of 98.5% of the added protein. In-run replicate precision (CV) of the method was 1.45%; day-to-day precision was 2.58%.Entities: