Literature DB >> 11572753

Tracheal-innominate artery fistula caused by the endotracheal tube tip: case report and investigation of a fatal complication of prolonged intubation.

M Siobal1, R H Kallet, R Kraemer, E Jonson, D Lemons, D Young, A R Campbell, W Schecter, J Tang.   

Abstract

CASE REPORT: A patient with extensive burns was intubated with an 8.0 mm internal diameter endotracheal tube (ETT) equipped with a subglottic suction port (Mallinckrodt HiLo Evac). The ETT was secured to a left upper molar with wire sutures throughout the hospitalization course to ensure airway stability. On the 40th day of intubation, the patient exsanguinated and died from a tracheo-innominate artery fistula. Postmortem examination revealed a 1 cm lesion of the left anterior tracheal wall at the position of the ETT tip. The prolonged stationary position of the ETT was considered the primary factor responsible for the fistula. Yet tracheo-innominate artery fistula normally is associated with high cuff pressures rather than with the tube tip. The special ETT construction required for the subglottic suction feature was suspected to have increased tube rigidity and may have played a contributory role.
METHODS: The rigidity of the Mallinckrodt HiLo Evac was measured with a mechanical model and compared to 5 other commercially-available ETTs. Rigidity was expressed as the force generated by the ETT tip when the tube curvature was altered by 5 cm and 10 cm of flexion from its resting position.
RESULTS: The mean force exerted by the Mallinckrodt HiLo Evac was 10.1 +/- 2.8 g at 5 cm of flexion and 17.7 +/- 5.1 g at 10 cm of flexion. This was significantly greater than all other ETT brands tested (by one-way analysis of variance and Student-Newman-Kuels test, p < 0.05).
CONCLUSION: This case of fatal tracheo-innominate artery fistula formation associated with an ETT tip was unusual because of the extended duration of endotracheal intubation and the complexity of the patient's airway management problems. Our data suggest that the higher rigidity of the HiLo Evac ETT may have contributed to fistula development at the tube tip. However, we do not believe that the higher rigidity of the HiLo Evac ETT necessarily poses any greater risk than other ETTs under normal circumstances, in which the tube tip is not fixed in a stationary position for an extended period.

Entities:  

Mesh:

Year:  2001        PMID: 11572753

Source DB:  PubMed          Journal:  Respir Care        ISSN: 0020-1324            Impact factor:   2.258


  2 in total

1.  Randomized Pilot Trial of Two Modified Endotracheal Tubes To Prevent Ventilator-associated Pneumonia.

Authors:  Steven Deem; David Yanez; Laura Sissons-Ross; Jo Ann Elrod Broeckel; Stephen Daniel; Miriam Treggiari
Journal:  Ann Am Thorac Soc       Date:  2016-01

Review 2.  Updating the evidence-base for suctioning adult patients: a systematic review.

Authors:  Tom J Overend; Cathy M Anderson; Dina Brooks; Lisa Cicutto; Michael Keim; Debra McAuslan; Mika Nonoyama
Journal:  Can Respir J       Date:  2009 May-Jun       Impact factor: 2.409

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.