Literature DB >> 11566250

Measurement of single kidney contrast media clearance by multiphasic spiral computed tomography: preliminary results.

N Hackstein1, M F Puille, B H Bak, O Scharwat, W S Rau.   

Abstract

OBJECTIVE: We present preliminary results of a new method (hereinafter called 'CT-clearance') to measure single kidney contrast media clearance by performing multiphasic helical CT of the kidneys. CT-clearance was calculated according to an extension of the Patlak-Plot. In contrast to prior investigators, who repeatedly measured a single slice, this method makes it possible to calculate single kidney clearance from at least three spiral CTs, utilizing the whole kidney volume.
METHODS: Spiral CT of the kidneys was performed unenhanced and about 30 and 100 s after administration of about 120 ml iopromide. Sum-density of the whole kidneys and aortic density was calculated from this data. Using this data, renal clearance of contrast media was calculated by CT-clearance in 29 patients. As reference, Serum-clearance was calculated in 24 patients by application of a modified one-exponential slope model. Information on the relative kidney function was gained by renal scintigraphy with Tc99m-MAG-3 or Tc99m-DMSA in 29 patients.
RESULTS: Linear regression analysis revealed a correlation coefficient of CT-clearance with Serum-clearance of r=0.78 with Cl (CT) [ml/min]=22.2+1.03 * Cl (serum), n=24. Linear regression of the relative kidney function (rkf) of the right kidney calculated by CT-clearance compared to scintigraphy results provided a correlation coefficient r=0.89 with rkf(CT)[%]=18.6+0.58 * rkf(scintigraphy), n=29.
CONCLUSION: The obtained results of contrast media clearance measured by CT-clearance are in the physiological range of the parameter. Future studies should be performed to improve the methodology with the aim of higher accuracy. More specifically, better determination of the aortic density curve might improve the accuracy.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11566250     DOI: 10.1016/s0720-048x(01)00323-0

Source DB:  PubMed          Journal:  Eur J Radiol        ISSN: 0720-048X            Impact factor:   3.528


  3 in total

1.  Determination of Single-Kidney Glomerular Filtration Rate in Human Subjects by Using CT.

Authors:  Soon Hyo Kwon; Ahmed Saad; Sandra M Herrmann; Stephen C Textor; Lilach O Lerman
Journal:  Radiology       Date:  2015-04-06       Impact factor: 11.105

2.  Determination of single-kidney glomerular filtration rate (GFR) with CT urography versus renal dynamic imaging Gates method.

Authors:  Shan You; XianWu Ma; ChangZhu Zhang; Qiang Li; WenWei Shi; Jing Zhang; XiaoDong Yuan
Journal:  Eur Radiol       Date:  2017-10-02       Impact factor: 5.315

3.  Measurement of Glomerular Filtration Rate Using Multiphasic Computed Tomography in Patients With Unilateral Renal Tumors: A Feasibility Study.

Authors:  Tingting Wang; Yi Xu; Wangyan Liu; Pengfei Shao; Qiang Lv; Guanyu Yang; Lijun Tang
Journal:  Front Physiol       Date:  2019-09-19       Impact factor: 4.566