Is the GraftConnector a valid alternative to running suture in end-to-side coronary arteries anastomoses?

P Tozzi1, J O Solem, D Boumzebra, A Mucciolo, C Y Genton, P Chaubert, L K von Segesser.   

Abstract

BACKGROUND: An animal study was carried out to compare long-term patency rates of coronary anastomoses performed with the GraftConnector versus running suture technique.
METHODS: 10 sheep, 45 to 55 kg, underwent off-pump coronary artery bypass grafting (right internal mammary artery to left anterior descending artery). In 5 animals, the anastomosis was performed with a GraftConnector and in 5 animals with 7-0 running suture. Intraoperative fluoroscopy and a fluoroscopic control at 6 months were performed. After 6 months, the animals were sacrificed and the anastomoses were examined histologically.
RESULTS: All animals survived at 6 months with 100% anastomosis patency rates in both groups. In the GraftConnector group, the anastomosis diameter at 6 months fluoroscopy was 118% of native left anterior descending artery versus 97% of the control group. Luminal anastomotic width at histology was 1.7 +/- 0.2 mm in the device group versus 1.6 +/- 0.1 mm in the control group. Mean intimal hyperplasia thickness was 0.21 +/- 0.1 mm in the device group versus 0.01 mm in the control group.
CONCLUSIONS: The GraftConnector provides a consistent and reproducible coronary artery anastomosis and reduces technical demand and manual dexterity in coronary operations. Long-term results demonstrate that off-pump coronary artery bypass grafting performed with the GraftConnector had the same patency rate and luminal width as those performed with running suture.

Entities:  

Mesh:

Year:  2001        PMID: 11565736     DOI: 10.1016/s0003-4975(01)02953-8

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  2 in total

1.  Internal thoracic artery to left anterior descending anastomosis performed on the beating heart by endoscopic robot assistance using a new distal connector.

Authors:  Inderjit S Gill; Hironori Izutani
Journal:  Can J Surg       Date:  2006-02       Impact factor: 2.089

2.  The vascular connector, design of a new device for sutureless vascular anastomosis.

Authors:  Lulzim Vokrri; Xhavit Krasniqi; Arsim Qavdarbasha; Nexhmi Hyseni; Philippe Cinquin; Paolo Porcu; Carmine Sessa
Journal:  Ann Surg Innov Res       Date:  2014-11-22
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.