Is the use of albumin in colloid prime solution of cardiopulmonary bypass circuit justified?

R H Boks1, L A van Herwerden, J J Takkenberg, W van Oeveren, Y J Gu, M J Wijers, A J Bogers.   

Abstract

BACKGROUND: Albumin in the priming solution precoats the surface of the cardiopulmonary bypass circuit, supposedly causing delayed adsorption of fibrinogen and reduced activation and adhesion of platelets. This action may result in lower transoxygenator resistance. Because our institution uses a colloidal prime solution (Gelofusine), questions were raised about the value of albumin in the prime solution. We decided to focus on the clinical effects of transoxygenator resistance.
METHODS: Sixty adults undergoing elective cardiac operations were randomly divided into three groups: a group with 20-g albumin (n = 20), a group with 2-g albumin (n = 20), and a group with no albumin (n = 20) in the 1,600-mL colloidal prime. Patients older than 75 years and patients with a preoperative serum albumin level of 30 g/L or less were excluded. The transoxygenator resistance was measured throughout cardiopulmonary bypass. Beta-thromboglobulin levels were used to study contact activation of platelets. Measures of prothrombin F1,2 fragments were used as a marker of thrombin generation. Body surface area, age, preoperative albumin, hematocrit, hemoglobin, fibrinogen, platelet count, and colloid osmotic pressure levels were compared between groups.
RESULTS: Base line characteristics and chosen control measurements were similar for all three populations. When comparing the observed transoxygenator resistance among the three different groups, no significant differences were noted. Prothrombin F1.2 fragments remained low for all the groups without significant differences. In the no-albumin group the level of beta-thromboglobulin appeared to be higher, but the difference was not statistically significant.
CONCLUSIONS: Addition of albumin to prime solution in a cardiopulmonary bypass circuit that already contains colloids does not affect the transoxygenator resistance of the COBE Duo flat sheet oxygenator and does not affect prothrombin F1.2 and beta-thromboglobulin levels. Therefore additional costs for the albumin are not justified. Measurement of transoxygenator resistance is a reliable, simple method to determine the effects of a prime solution on the oxygenator surface in vivo.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11565669     DOI: 10.1016/s0003-4975(01)02816-8

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  4 in total

1.  Getting it right: optimizing the patient and technique for the procedure.

Authors:  Alfred H Stammers
Journal:  J Extra Corpor Technol       Date:  2009-12

2.  The quantitative and qualitative responses of platelets in pediatric patients undergoing cardiopulmonary bypass surgery.

Authors:  Vera Ignjatovic; Jenny Than; Robyn Summerhayes; Fiona Newall; Steve Horton; Andrew Cochrane; Paul Monagle
Journal:  Pediatr Cardiol       Date:  2011-08-02       Impact factor: 1.655

3.  Effects of crystalloid and colloid priming strategies for cardiopulmonary bypass on colloid oncotic pressure and haemostasis: a meta-analysis.

Authors:  Anne Maria Beukers; Jamy Adriana Catharina de Ruijter; Stephan Alexander Loer; Alexander Vonk; Carolien Suzanna Enna Bulte
Journal:  Interact Cardiovasc Thorac Surg       Date:  2022-08-03

4.  Evaluation of Different Dialyzers and the Impact of Predialysis Albumin Priming in Intermittent Hemodialysis With Reduced Anticoagulation.

Authors:  Floris Vanommeslaeghe; Filip De Somer; Iván Josipovic; Matthieu Boone; Wim Van Biesen; Sunny Eloot
Journal:  Kidney Int Rep       Date:  2019-07-24
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.