Detector sensitivity influences blood flow sampling in scanning laser Doppler flowmetry.

S L Hosking1, S Embleton, L Kagemann, A Chabra, C Jonescu-Cuypers, A Harris.   

Abstract

PURPOSE: To establish the effect of photodiode sensitivity on the DC (brightness) value and the resultant blood flow measurements of retina and rim tissue using a scanning laser Doppler flowmeter (SLDF).
METHODS: The sample consisted of one eye of each of 15 healthy subjects (mean age 27.8 +/- 6.1 years). Using the Heidelberg Retina Flowmeter (HRF), three 10-deg images of the superior temporal retina and three further images of the superior temporal rim were acquired for each of five DC bands: band 1: 30-70; band 2: 70-110; band 3: 110-150; band 4 150-190; band 5: 190-230. Retinal blood volume, flow and velocity were determined for each image using a 10 x 10 pixel square grid located at a predetermined location on the retina and rim for each subject. Following image acquisition, the DC values corresponding to each pre-assigned retinal or rim location were determined. The mean and standard deviation were determined for the blood flow parameters within each DC band for each subject in both locations. Analysis of variance was used to identify significant change in the data as a function of the DC value (P<0.05).
RESULTS: Analysis of variance revealed that retinal blood flow measures acquired within DC band 5 resulted in significantly lower measures of blood flow and velocity (P=0.035 and P=0.049 respectively) than at lower DC values. Band 5 values of flow, volume and velocity in the neuroretinal rim were also significantly low (P=0.016, P= 0.003 and P=0.026 respectively). Peak neuroretinal rim blood flow was recorded when the DC value was between 70 and 110. For blood flow measurement at the retina and neuroretinal rim the DC value should not exceed 190.
CONCLUSION: Photodiode sensitivity as indicated by the DC value affects measurements of ocular blood flow using the HRF.

Entities:  

Mesh:

Year:  2001        PMID: 11561787     DOI: 10.1007/s004170100283

Source DB:  PubMed          Journal:  Graefes Arch Clin Exp Ophthalmol        ISSN: 0721-832X            Impact factor:   3.117


  5 in total

1.  The effect of image alignment on capillary blood flow measurement of the neuroretinal rim using the Heidelberg retina flowmeter.

Authors:  M Sehi; J G Flanagan
Journal:  Br J Ophthalmol       Date:  2004-02       Impact factor: 4.638

2.  Reproducibility of circadian retinal and optic nerve head blood flow measurements by Heidelberg retina flowmetry.

Authors:  C P Jonescu-Cuypers; A Harris; K U Bartz-Schmidt; L Kagemann; A S Boros; U E Heimann; B H Lenz; R-D Hilgers; G K Krieglstein
Journal:  Br J Ophthalmol       Date:  2004-03       Impact factor: 4.638

3.  Ocular blood flow measurements in healthy human myopic eyes.

Authors:  Alexandra Benavente-Pérez; Sarah L Hosking; Nicola S Logan; David C Broadway
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2010-05-26       Impact factor: 3.117

4.  Impact of simulated light scatter on scanning laser Doppler flowmetry.

Authors:  S T Venkataraman; C Hudson; E Harvey; J G Flanagan
Journal:  Br J Ophthalmol       Date:  2005-09       Impact factor: 4.638

5.  Strategies for reducing variance in laser Doppler flowmetry measurements.

Authors:  Berthold Pemp; Noemi Maar; Guenther Weigert; Alexandra Luksch; Hemma Resch; Gerhard Garhofer; Selim Orgul; Leopold Schmetterer
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2008-09-10       Impact factor: 3.117

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.