The use of computed tomography when comparing nickel-titanium and stainless steel files during preparation of simulated curved canals.

Y Garip1, M Günday.   

Abstract

AIM: The aim of this study was to compare canal preparations completed with Hedstrom and K-files of ISO size 15 - 40, made of nickel-titanium (Ni-Ti) and stainless steel (SS).
METHODOLOGY: Eighty simulated canals with 200 and 30 degrees C curvatures were prepared using the step-back technique and quarter turn/pull instrument manipulation. Middle and apical level canal sections were taken using computed tomography.
RESULTS: No significant difference was found between any of the file types at either level with respect to canal curvature (20 degrees or 30 degrees). At the middle level, the stainless steel files caused more enlargement toward the inner part, compared to nickel-titanium files. At the apical level, nickel-titanium canal files caused more enlargement toward the inner part, whereas more outward enlargement was caused by stainless steel instruments. No significant difference could be observed at the middle level (P > 0.05) related to the enlargement toward the outer side of the canal curvature. Transportation at both levels was significantly less (P < 0.001) for the Ni-Ti files than the SS ones. Centring ratios of the file types at the middle level were low, but not significantly different (P > 0.05), whereas at the apical level the centring ratios were significantly higher for the Ni-Ti files (P < 0.001).
CONCLUSIONS: Ni-Ti instruments produced preparations with adequate enlargement, less transportation, and a better centring ratio.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11556512     DOI: 10.1046/j.1365-2591.2001.00416.x

Source DB:  PubMed          Journal:  Int Endod J        ISSN: 0143-2885            Impact factor:   5.264


  13 in total

1.  A comparison of the shaping ability of reciprocating NiTi instruments in simulated curved canals.

Authors:  Young-Sil Yoo; Yong-Bum Cho
Journal:  Restor Dent Endod       Date:  2012-11-21

2.  CT evaluation of canal preparation using rotary and hand NI-TI instruments: An in vitro study.

Authors:  Shruthi Nagaraja; B V Sreenivasa Murthy
Journal:  J Conserv Dent       Date:  2010-01

3.  Evaluation of Forces Generated on Three Different Rotary File Systems in Apical Third of Root Canal using Finite Element Analysis.

Authors:  Ashish Medha; Suvarna Patil; Upendra Hoshing; Siddhesh Bandekar
Journal:  J Clin Diagn Res       Date:  2013-12-09

4.  Cyclic fatigue resistance of four nickel-titanium rotary instruments: a comparative study.

Authors:  Eugenio Pedullà; Gianluca Plotino; Nicola Maria Grande; Alfio Pappalardo; Ernesto Rapisarda
Journal:  Ann Stomatol (Roma)       Date:  2012-08-09

5.  An ex vivo comparative analysis on shaping ability of four NiTi rotary endodontic instruments using spiral computed tomography.

Authors:  Nitin Maitin; D Arunagiri; Dexter Brave; Shipra Nangalia Maitin; Sandeep Kaushik; Saumya Roy
Journal:  J Conserv Dent       Date:  2013-05

6.  The effect of a manual instrumentation technique on five types of premolar root canal geometry assessed by microcomputed tomography and three-dimensional reconstruction.

Authors:  Ke-Zeng Li; Yuan Gao; Ru Zhang; Tao Hu; Bin Guo
Journal:  BMC Med Imaging       Date:  2011-06-15       Impact factor: 1.930

7.  Taper preparation variability compared to current taper standards using computed tomography.

Authors:  Richard Gergi; Joe Abou Rjeily; Nada Osta; Joseph Sader; Alfred Naaman
Journal:  Int J Dent       Date:  2012-02-29

8.  Centering ability and dentin removal of rotary systems in curved root canals.

Authors:  Saeed Moradi; Ali Talati; Ali Monajem Zadeh
Journal:  Iran Endod J       Date:  2009-07-06

9.  Comparison of removed dentin thickness with hand and rotary instruments.

Authors:  Shahriar Shahriari; Hasan Abedi; Mahdi Hashemi; Seyed Mohsen Jalalzadeh
Journal:  Iran Endod J       Date:  2009-04-17

10.  Cone-Beam Computed Tomography Comparison of Canal Transportation after Preparation with BioRaCe and Mtwo Rotary Instruments and Hand K-Flexofiles.

Authors:  Hadi Mokhtari; Mahdi Niknami; Aydin Sohrabi; Ehsan Habibivand; Hamid Reza Mokhtari Zonouzi; Saeed Rahimi; Vahid Zand
Journal:  Iran Endod J       Date:  2014-07-05
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.