Abstract
A new proportional assist ventilation (PAV) method using a proportional solenoid valve (PSV) to control air supply to patients suffering from respiratory disabilities, was studied. The outlet flow and pressure from the proportional solenoid valve at various air supply pressures were tested and proven to be suitable for pressure and flow control in a PAV system. In vitro tests using a breathing simulator, which has been proven to possess the general characteristics of human respiratory system in spontaneous breathing tests, were conducted and the results demonstrated the viability of this PAV system in normalizing the breathing patterns of patients with abnormally high resistances and elastances as well as neuromuscular weaknesses. With a back-up safety mechanism incorporated in the control program, pressure "run-away" can be effectively prevented and safe operation of the system can be guaranteed.Entities:
Mesh:
Year: 2001 PMID: 11551814 DOI: 10.1016/s1350-4533(01)00058-3
Source DB: PubMed Journal: Med Eng Phys ISSN: 1350-4533 Impact factor: 2.242