Experimental upgrades of membrane introduction mass spectrometry for water and air analysis.

P Bocchini1, R Pozzi, C Andalò, G C Galletti.   

Abstract

Some improvements to the membrane introduction mass spectrometry (MIMS) technique, resulting in low-ppt detection limits for volatile organohalogen compounds (CX) in water (namely, chloroform, bromoform, bromodichloromethane, chlorodibromomethane, tetrachloroethylene, trichloroethylene, 1,1,1-trichloroethane, and carbon tetrachloride) and low-microgram per cubic meter detection limits for benzene, toluene, ethylbenzene, and xylenes (BTEX) in gaseous samples, are shown. A static MIMS configuration was compared to a dynamic one, the former requiring longer time to obtain the analytical response. A cryotrapping preconcentration step is introduced and linearity of response, mixture effects, and detection limits are presented. The instrumental setup consists of a hollow fiber silicone membrane, a water or air container, a cryofocusing trap based on Tenax adsorbent, a Peltier cell, and a Varian ion trap benchtop mass spectrometer is described. This instrumental setup, which we named membrane extraction trap focusing mass spectrometry, allowed the detection of CX in water at a concentration as low as 8 ppt and of benzene in air at 0.1 microg/m3. The whole assembly shows great potential for on-site routine monitoring of drinking water resources and urban and indoor air under current EU and Italian regulations.

Entities:  

Year:  2001        PMID: 11534703     DOI: 10.1021/ac010249e

Source DB:  PubMed          Journal:  Anal Chem        ISSN: 0003-2700            Impact factor:   6.986


  1 in total

1.  An enzyme derivatized polydimethylsiloxane (PDMS) membrane for use in membrane introduction mass spectrometry (MIMS).

Authors:  A Skye Creba; Alexandra N E Weissfloch; Erik T Krogh; Chris G Gill
Journal:  J Am Soc Mass Spectrom       Date:  2007-03-28       Impact factor: 3.109

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.