Literature DB >> 11525551

Precision of three topography instruments in keratoconus subjects.

T T McMahon1, R J Anderson, C E Joslin, G A Rosas.   

Abstract

PURPOSE: To determine the test-retest reliability of three popular and commercially available Placido-ring videokeratography instruments in subjects with keratoconus.
METHODS: Nine subjects (16 eyes) with keratoconus of varying degrees of severity had up to four images per eye generated, in random order, from the EyeSys Model II, Dicon CT 200 and the Keratron Corneal Analyzer. Test-retest analyses for the images were sampled at four locations: 1.5 mm nasal, inferior, temporal, and superior from center. The average standard deviation of all points was used to determine the short-term variability of the measurements.
RESULTS: The short-term variability (in diopters) of the Dicon, EyeSys, and Keratron was 0.61 to 3.31 D, 0.94 to 1.51 D, and 0.58 to 2.85 D, respectively, for axial distance maps and 1.07 to 6.82 D, 0.79 to 1.77 D, and 1.23 to 3.03 D for tangential curvature maps.
CONCLUSION: Results support the notion of a loss in repeatability for all three instruments when corneal irregularity is present, which reduces test-retest reliability.

Mesh:

Year:  2001        PMID: 11525551     DOI: 10.1097/00006324-200108000-00012

Source DB:  PubMed          Journal:  Optom Vis Sci        ISSN: 1040-5488            Impact factor:   1.973


  5 in total

1.  Validity of autorefractor based screening method for irregular astigmatism compared to the corneal topography- a cross sectional study.

Authors:  Alicia Galindo-Ferreiro; Julita De Miguel-Gutierrez; Manuel González-Sagrado; Alberto Galvez-Ruiz; Rajiv Khandekar; Silvana Schellini; Julio Galindo-Alonso
Journal:  Int J Ophthalmol       Date:  2017-09-18       Impact factor: 1.779

2.  Difficulties in choosing the right intraocular lens in a previously vitrectomized patient - the role of the tear film.

Authors:  Monica Mălăescu; Bogdana Tăbăcaru; Horia Tudor Stanca
Journal:  Rom J Ophthalmol       Date:  2022 Jan-Mar

3.  Effects of multicurve RGP contact lens use on topographic changes in keratoconus.

Authors:  Joon Seo Hwang; Jin Hak Lee; Won Ryang Wee; Mee Kum Kim
Journal:  Korean J Ophthalmol       Date:  2010-08-03

4.  Cataract Surgery in Patients with Keratoconus: Pearls and Pitfalls.

Authors:  F Aiello; Q J Nasser; C Nucci; R I Angunawela; Z Gatzioufas; V Maurino
Journal:  Open Ophthalmol J       Date:  2017-07-31

Review 5.  Patient selection for corneal collagen cross-linking: an updated review.

Authors:  Virgilio Galvis; Alejandro Tello; Alvaro I Ortiz; Luis C Escaf
Journal:  Clin Ophthalmol       Date:  2017-04-07
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.