Literature DB >> 11524645

Performance and error analysis of automated external defibrillator use in the out-of-hospital setting.

R D Macdonald1, J M Swanson, J L Mottley, C Weinstein.   

Abstract

STUDY
OBJECTIVE: We determined whether automated external defibrillators (AEDs) can meet the American Heart Association performance criteria to detect and shock unstable cardiac rhythms (ventricular fibrillation [VF], ventricular tachycardia [VT]) in the setting of an out-of-hospital cardiac arrest.
METHODS: AED performance was reviewed for cardiac arrests occurring between January 1, 1995, and December 31, 1997. After every cardiac arrest, data regarding each rhythm analyzed and subsequent response (shock or no shock) were downloaded from the AED memory module. The study paramedic and study physician independently reviewed each case and interpreted cardiac rhythms from downloaded AED data. The emergency medical services medical director resolved all discrepancies in a blinded manner. All cases of out-of-hospital cardiac arrest in which an AED was turned on and a rhythm analyzed were included. The primary objective was the correct identification and defibrillation of VF or VT. Sensitivity, specificity, and predictive values with 95% confidence intervals (CIs) were calculated. Sources of error in AED rhythm management are also described.
RESULTS: A total of 3,448 AED rhythms were available for interpretation. Sensitivity and specificity for appropriate AED management of a shockable (VF or VT) rhythm were 81.0% (95% CI 77.9% to 83.8%) and 99.9% (95% CI 99.7% to 100%), respectively. Positive and negative predictive values were 99.6% (95% CI 98.7% to 99.9%) and 95.5% (95% CI 94.7% to 96.2%), respectively. There were 132 errors associated with AED management. Two errors resulted in delivery of an inappropriate shock. In the remaining 130 errors, a shockable rhythm was not shocked. Fifty-five (42.3%) errors were AED dependent, 70 (53.9%) were operator dependent, and 5 (3.9%) were unclassified.
CONCLUSION: The AED had high specificity and moderately high sensitivity in detecting and shocking unstable cardiac rhythms in the out-of-hospital setting. Few cardiac rhythms were mismanaged by the AED. Elimination of operator-dependent errors could increase AED sensitivity.

Entities:  

Mesh:

Year:  2001        PMID: 11524645     DOI: 10.1067/mem.2001.117953

Source DB:  PubMed          Journal:  Ann Emerg Med        ISSN: 0196-0644            Impact factor:   5.721


  6 in total

1.  Must we always follow the instructions of automated external defibrillators?

Authors:  Nicolas Paleiron; Anne Pegorie; Ba Vinh Nguyen; Christophe Giacardi; Diane Commandeur; Marc Danguy des Déserts; Medhi Ould-Ahmed
Journal:  Intensive Care Med       Date:  2009-12-15       Impact factor: 17.440

2.  Ventricular tachyarrhythmias after cardiac arrest in public versus at home.

Authors:  Myron L Weisfeldt; Siobhan Everson-Stewart; Colleen Sitlani; Thomas Rea; Tom P Aufderheide; Dianne L Atkins; Blair Bigham; Steven C Brooks; Christopher Foerster; Randal Gray; Joseph P Ornato; Judy Powell; Peter J Kudenchuk; Laurie J Morrison
Journal:  N Engl J Med       Date:  2011-01-27       Impact factor: 91.245

3.  Effect of real-time feedback during cardiopulmonary resuscitation outside hospital: prospective, cluster-randomised trial.

Authors:  David Hostler; Siobhan Everson-Stewart; Thomas D Rea; Ian G Stiell; Clifton W Callaway; Peter J Kudenchuk; Gena K Sears; Scott S Emerson; Graham Nichol
Journal:  BMJ       Date:  2011-02-04

4.  Diagnostic Accuracy of Commercially Available Automated External Defibrillators.

Authors:  Takahiko Nishiyama; Ako Nishiyama; Masachika Negishi; Shin Kashimura; Yoshinori Katsumata; Takehiro Kimura; Nobuhiro Nishiyama; Yoko Tanimoto; Yoshiyasu Aizawa; Hideo Mitamura; Keiichi Fukuda; Seiji Takatsuki
Journal:  J Am Heart Assoc       Date:  2015-12-01       Impact factor: 5.501

5.  Erroneous shock by an AED: Importance of obtaining AED tracing to prevent inappropriate ICD implantation.

Authors:  Brian J Cross; Mark S Link
Journal:  HeartRhythm Case Rep       Date:  2015-02-12

6.  Clinical outcomes and safety of passive leg raising in out-of-hospital cardiac arrest: a randomized controlled trial.

Authors:  Youcef Azeli; Alfredo Bardají; Eneko Barbería; Vanesa Lopez-Madrid; Jordi Bladé-Creixenti; Laura Fernández-Sender; Gil Bonet; Elena Rica; Susana Álvarez; Alberto Fernández; Christer Axelsson; Maria F Jiménez-Herrera
Journal:  Crit Care       Date:  2021-05-25       Impact factor: 9.097

  6 in total