Evaluation of Aution Max AX-4280 automated urine test-strip analyser.

Z Zaman1, S Roggeman, P Cappelletti, G Ferrai, M Buxeda, N Barba.   

Abstract

Aution Max AX-4280, an automated urine test-strip analyser, was evaluated in three centres. Method comparison, imprecision, carry-over, linearity, detection limit and drift studies were performed for glucose, protein, blood and leukocytes using Uriflet S 9UB strips. These strips enable measurement of pH, glucose, protein, blood, leukocytes, ketones, bilirubin, urobilinogen and nitrite. Specific gravity is determined by the refractive index method. Within-run and between-day imprecision, assessed using pooled urines and quality control materials, were good. No drift over 24 h or sample carry-over was observed. Method comparison with quantitative methods for glucose, protein and specific gravity yielded good correlations. Ascorbate negatively interfered with haemoglobin, glucose and nitrite measurements. Acetylsalicylic acid lowered pH, the effect being greatest when protein was absent. During the assessment period no malfunction or breakdown was reported. The Aution Max is easy to use and needs minimal maintenance.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11522115     DOI: 10.1515/CCLM.2001.106

Source DB:  PubMed          Journal:  Clin Chem Lab Med        ISSN: 1434-6621            Impact factor:   3.694


  3 in total

1.  Analytical performance, agreement and user-friendliness of six point-of-care testing urine analysers for urinary tract infection in general practice.

Authors:  Marjolein J C Schot; Sanne van Delft; Antoinette M J Kooijman-Buiting; Niek J de Wit; Rogier M Hopstaken
Journal:  BMJ Open       Date:  2015-05-18       Impact factor: 2.692

2.  Comparison of the changes in the factors associated with the renal prognosis of non-elderly and elderly subjects treated with empagliflozin- a retrospective observation study in Japanese patients with type 2 diabetes.

Authors:  Hiroyuki Ito; Suzuko Matsumoto; Takuma Izutsu; Eiji Kusano; Shinya Nishio; Shinichi Antoku; Tomoko Yamasaki; Toshiko Mori; Michiko Togane; Shigenori Ando; Emiko Tsugami
Journal:  Diabetes Metab Syndr Obes       Date:  2019-09-09       Impact factor: 3.168

3.  Prospective, observational study comparing automated and visual point-of-care urinalysis in general practice.

Authors:  Sanne van Delft; Annelijn Goedhart; Mark Spigt; Bart van Pinxteren; Niek de Wit; Rogier Hopstaken
Journal:  BMJ Open       Date:  2016-08-08       Impact factor: 2.692

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.