We examine the effects of aberrations induced by a refractive index mismatch on the signal level and resolution of single-photon (1-p) and two-photon (2-p), conventional and confocal scanning microscopes. In particular, we consider the aberrations introduced by an interface between oil/glass and water. Resolution is defined in terms of enclosed fluorescence, rather than full-width half-maximum, revealing more useful information for heavily aberrated point spread functions (PSFs). It is shown that, at large focusing depths, the resolution of 2-p conventional and 1-p confocal microscopes are almost identical. The benefits of aberration correction are examined by removing Zernike aberration modes. With aberration correction, the best resolution is found for 1-p confocal and 2-p confocal modes. An approximation based upon geometrical optics is also introduced which shows that the axial resolution of heavily aberrated PSFs is roughly proportional to focusing depth.Entities: