| Literature DB >> 11513024 |
S D Metzler1, J E Bowsher, M F Smith, R J Jaszczak.
Abstract
Pinhole collimators are widely used to image small organs and animals. The sensitivity of knife-edge pinhole collimators has been previously estimated using an "effective diameter" formulation and experimentally described using a sin(x) theta fit, where theta is the angle between the line segment from the center of the aperture to the photon source and its projection onto the plane of the aperture. An analytic form of the sensitivity of the pinhole collimator is derived in this paper. A numerical formula for predicting the sin(x) theta form of the sensitivity is calculated from the analytic form. Experimental data are compared with the theoretical estimate and the sin(x) theta prediction. The agreement is excellent.Entities:
Mesh:
Year: 2001 PMID: 11513024 DOI: 10.1109/42.938241
Source DB: PubMed Journal: IEEE Trans Med Imaging ISSN: 0278-0062 Impact factor: 10.048