Abstract

PURPOSE: To compare the accuracy of three camera-assisted methods for the measurement of Tc-99m mercaptoacetyltriglycine (MAG3) clearance.
MATERIALS AND METHODS: MAG3 renal scintigraphy was performed in 21 adults with different degrees of renal function. Posterior views were obtained that included the heart and the kidneys in the camera field of view. The syringe for injection was imaged before and after injection. Blood samples were drawn 24 and 43 minutes after injection and plasma radioactivity was measured. Three camera-assisted techniques to measure MAG3 clearance were tested: 1) Biexponential fitting of the left ventricular curve, normalized to plasma activity at 24 minutes; 2) calculation of clearance by relating the integral of the plasma curve (normalized to plasma activity) to the kidney activity using the Rutland-Patlak space method; and 3) a regression equation measuring clearance from the percentage of the injected dose accumulating in the kidneys during the 1- to 2.5-minute period. The camera-assisted clearances were compared with the single-sample MAG3 clearances calculated using the Russell equation. Linear regression analysis was used to measure the correlation between the camera-based methods and the single-sample techniques.
RESULTS: Correlation with r > 0.900 was found for all three techniques. The difference in correlation coefficients between the three methods was not significant; however, the regression line of method 3 was significantly closer to the line of identity (P = 0.005).
CONCLUSION: Method 3 most closely fits the line of identity and is probably the most practical because no blood sample is needed.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11507290     DOI: 10.1097/00003072-200109000-00001

Source DB:  PubMed          Journal:  Clin Nucl Med        ISSN: 0363-9762            Impact factor:   7.794


  3 in total

1.  Assessment of effective renal plasma flow, enzymuria, and cytokine release in healthy volunteers receiving a single dose of amphotericin B desoxycholate.

Authors:  Manjunath P Pai; Jeffrey P Norenberg; Robert A Telepak; David S Sidney; Shu Yang
Journal:  Antimicrob Agents Chemother       Date:  2005-09       Impact factor: 5.191

Review 2.  Radionuclides in nephrourology, part 1: Radiopharmaceuticals, quality control, and quantitative indices.

Authors:  Andrew T Taylor
Journal:  J Nucl Med       Date:  2014-02-18       Impact factor: 10.057

3.  Decision support systems in diuresis renography.

Authors:  Andrew Taylor; Amita Manatunga; Ernest V Garcia
Journal:  Semin Nucl Med       Date:  2008-01       Impact factor: 4.446

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.