| Literature DB >> 11499528 |
Abstract
This paper discusses the single-unit recording characteristics of microelectrode arrays containing on-chip signal processing circuitry. Probes buffered using on-chip unity-gain operational amplifiers provide an output resistance of 200 ohm with an input-referred noise of 11-muV root-mean-square (rms) (100 Hz-10 kHz). Simultaneous in vivo recordings from single neurons using buffered and unbuffered (passive) iridium recording sites separated by less than 20 microm have shown that the use of on-chip circuitry does not significantly degrade system noise. Single-unit neural activity has also been studied using probes containing closed-loop preamplifiers having a voltage gain of 40 dB and a bandwidth of 13 kHz, and several input dc-baseline stabilization techniques have been evaluated. Low-noise in vivo recordings with a multiplexed probe have been demonstrated for the first time using an external asymmetrical clock running at 200 kHz. The multiplexed system adds less than 8-muV rms of noise to the recorded signals, suppressing the 5-V clock transitions to less than 2 ppm.Entities:
Mesh:
Year: 2001 PMID: 11499528 DOI: 10.1109/10.936367
Source DB: PubMed Journal: IEEE Trans Biomed Eng ISSN: 0018-9294 Impact factor: 4.538