Literature DB >> 11496268

Serious complications that require surgical interventions after endoluminal stent-graft placement for the treatment of infrarenal aortic aneurysms.

C Schlensak1, T Doenst, M Hauer, J Bitu-Moreno, P Uhrmeister, G Spillner, F Beyersdorf.   

Abstract

BACKGROUND: Endoluminal stent-graft placement for the treatment of infrarenal abdominal aortic aneurysms has gained widespread acceptance because it is associated with lower peri-interventional morbidity than conventional transabdominal surgery. In this study the long-term morbidity of the procedure was evaluated. METHODS AND
RESULTS: Between September 1994 and December 1998, 150 patients (age, 69.6 +/- 8.49 years; 142 men, 8 women) with abdominal aortic aneurysms were treated with the placement of an intravascular nitinol stent-graft (Stentor [55] and Vanguard-System [95]); there were eight tubular grafts and 142 bifurcated grafts. Initial placement of the stent-graft was successful in 144 patients. Mean follow-up was 49 +/- 25 months. In 13.3% of stent-graft placements we encountered the following complications: 4 cases of migration or dislocation of the prosthesis (30.5 +/- 7.4 months after placement); 2 ruptures of the aorta (26.7 and 15.0 months after placement); 3 recurrent thromboses of the stent-graft (25.5 +/- 5.3 months after placement); 3 endoleaks (27.5 +/- 15.7 months after placement); and 5 infections of the prosthesis (26.6 +/- 16.5 months after placement). There was no correlation between the complications and the type of stent used. All of these patients were treated by surgical replacement of the prosthesis with a Dacron graft.
CONCLUSIONS: (1) The results suggest that most complications are due to a continuation of the disease process leading to loosening of the prosthesis. (2) Explantation of the prosthesis and surgical repair is feasible but bears additional risks. (3) Because the onset of reperfusion of the excluded aneurysm cannot be predicted, all patients with infrarenal aortic stent-grafts require frequent computer tomographic follow-up. (4) Finally, the results call for further improvements in the design of the stent-graft.

Entities:  

Mesh:

Year:  2001        PMID: 11496268     DOI: 10.1067/mva.2001.116975

Source DB:  PubMed          Journal:  J Vasc Surg        ISSN: 0741-5214            Impact factor:   4.268


  2 in total

1.  Hybrid-repair of thoraco-abdominal or juxtarenal aortic aneurysm: what the radiologist should know.

Authors:  Tobias Krauss; Thomas Pfammatter; Dieter Mayer; Mario Lachat; Lukas Hechelhammer; Borut Marincek; Thomas Frauenfelder
Journal:  Eur Radiol       Date:  2009-09-30       Impact factor: 5.315

Review 2.  Aortoenteric fistula as a complication of open reconstruction and endovascular repair of abdominal aorta.

Authors:  Marek Tagowski; Hendryk Vieweg; Christian Wissgott; Reimer Andresen
Journal:  Radiol Res Pract       Date:  2014-09-14
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.