Air flow resistance of three heat and moisture exchanging filter designs under wet conditions: implications for patient safety.

N J Morgan-Hughes1, G H Mills, D Northwood.   

Abstract

Heat and moisture exchanging filters (HMEFs) can be blocked by secretions. We have studied HMEF performance under wet conditions to see which particular design features predispose to this complication. Dar Hygrobac-S (composite felt filter and cellulose exchanger), Dar Hygroster (composite pleated ceramic membrane and cellulose exchanger) and Pall BB22-15 (pleated ceramic membrane) HMEFs were tested. Saline retention, saline concealment, and changes in air flow resistance when wet were assessed. The cellulose exchanger in the composite Hygrobac-S and Hygroster retained saline, producing a 'tampon' effect, associated with bi-directional air flow resistances in excess of the international standard of a 5 cm H(2)O pressure drop at 60 litre min(-1) air flow. Furthermore, high air flow resistances occurred before free saline was apparent within the transparent filter housing. The pleat only BB22-15 showed a significant increase in expiratory air flow resistance, but only after the presence of saline was apparent. These data imply that composite HMEFs with cellulose exchangers are more likely to block or cause excessive work of breathing as a result of occult accumulation of patient secretions than pleat only HMEFs.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11493505     DOI: 10.1093/bja/87.2.289

Source DB:  PubMed          Journal:  Br J Anaesth        ISSN: 0007-0912            Impact factor:   9.166


  3 in total

1.  Evaluation the effect of breathing filters on end-tidal carbon dioxide during inferior abdominal surgery in infants and changes of tidal volume and respiratory rate needs for preventing of increasing end-tidal carbon dioxide.

Authors:  Parvin Sajedi; Mohsen Abooei; Amir Shafa; Mahboobeh Karbalaei; Atefeh Babaei
Journal:  J Res Med Sci       Date:  2016-11-07       Impact factor: 1.852

2.  Influence of a heat and moisture exchanger with a microbiological filter on measurements of maximal respiratory pressures and vital capacity in patients with COPD.

Authors:  Jeanette Janaina Jaber Lucato; Renata Cléia Claudino Barbosa; Patricia Salerno de Almeida Picanço; Thiago Marraccini Nogueira da Cunha; Renato Fraga Righetti
Journal:  J Bras Pneumol       Date:  2019-12-13       Impact factor: 2.624

3.  Ventilatory changes during the use of heat and moisture exchangers in patients submitted to mechanical ventilation with support pressure and adjustments in ventilation parameters to compensate for these possible changes: a self-controlled intervention study in humans.

Authors:  Jeanette Janaina Jaber Lucato; Thiago Marraccini Nogueira da Cunha; Aline Mela Dos Reis; Patricia Salerno de Almeida Picanço; Renata Cléia Claudino Barbosa; Joyce Liberali; Renato Fraga Righetti
Journal:  Rev Bras Ter Intensiva       Date:  2017 Apr-Jun
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.