| Literature DB >> 11489397 |
Abstract
A high-performance liquid chromatography isocratic procedure was developed for the assay of sertraline in drug substance and tablets. The chromatographic system consists of a RP-8 column (125 x 4 mm, 5 microm), a mobile phase composed of acetonitrile and sodium phosphate buffer, pH 5.5 (7:3), flow rate of 1.0 ml.min(-1) and UV detection at 270 nm. The method validation yielded good results. The coefficient of variation varied between 0.19 and 1.04% and accuracy of 99.18% was found. Calibration curve was linear between 0.5 and 2.5 mg x ml(-1); its correlation coefficient was 0.9999.Entities:
Mesh:
Substances:
Year: 2001 PMID: 11489397 DOI: 10.1016/s0731-7085(01)00425-3
Source DB: PubMed Journal: J Pharm Biomed Anal ISSN: 0731-7085 Impact factor: 3.935