L Gatzoulis1, R J Watson, L B Jordan, S D Pye, T Anderson, N Uren, D M Salter, K A Fox, W N McDicken.
Abstract
The aim of this work was to investigate the suitability of a novel forward-viewing intravascular ultrasound (IVUS) technique for three-dimensional imaging of severely stenosed or totally occluded vessels, where the conventional side-viewing IVUS systems are of limited use. A stiff 3.8 mm diameter forward-viewing catheter was manufactured to scan a 72 degrees sector ahead of its tip. Conical volume data were acquired by rotating the catheter over 180 degrees by means of a motorised mechanical system. Operating at 30 MHz, the catheter was integrated with an IVUS scanner and a radiofrequency data acquisition system. Postmortem carotid and femoral arteries were scanned in vitro. Correlation of the reconstructed images with histology demonstrated the ability of this forward-viewing IVUS system to visualise healthy lumens, bifurcations, thickened atherosclerotic walls and, most importantly, severe and complete vessel occlusions. A rotating-sector forward-viewing IVUS system is suitable for anatomical assessment of severely diseased vessels in three dimensions.Entities:
Mesh:
Year: 2001 PMID: 11476931 DOI: 10.1016/s0301-5629(01)00371-4
Source DB: PubMed Journal: Ultrasound Med Biol ISSN: 0301-5629 Impact factor: 2.998