Pre-clinical validation of a new partially cemented femoral prosthesis by synergetic use of numerical and experimental methods.

M Viceconti1, L Cristofolini, M Baleani, A Toni.   

Abstract

The present work reports the pre-clinical validation of an innovative partially cemented femoral prosthesis called cement-locked uncemented (CLU) prosthesis. The inventors of the device under investigation claimed that, when compared to a comparable fully cemented stem, the new stem would present various advantages. Two previous experimental studies confirmed that primary stability and stress shielding were comparable to those of cemented stems. Aim of the present study was to investigate if the remaining claims were confirmed as well. A complete finite element model of the bone-implant complex was created from CT data. The model was validated against in vitro measurements of bone surface strains as well as against primary stability measurements. The peak stresses predicted in the CLU cement mantle were not found significantly lower than those reported in other studies on fully cemented stems. However, once the cement inlet geometry is optimised and the associated stress risers are eliminated, the CLU cement mantle should be subjected to much lower stresses. The stress induced in the stems by both load cases was well below the fatigue limit of the Ti6Al4V alloy. Finite element models predicted for all load cases relative motion between cement and metal lower than 60 microm. This amplitude may be fully accommodated by elastic deformations of the cement micro-ridges. The experimental and numerical results showed the validity of the new fixation concept, although a further optimisation of the geometry of the cement pockets is needed in order to further reduce the stresses in the cement.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11470109     DOI: 10.1016/s0021-9290(01)00029-x

Source DB:  PubMed          Journal:  J Biomech        ISSN: 0021-9290            Impact factor:   2.712


  2 in total

1.  Outcome of hybrid stem fixation in osteoporotic female patients. A minimum five-year follow-up study.

Authors:  Francesco Traina; Enrico Tassinari; Federico Pilla; Marcello De Fine; Federico Biondi; Luca Cristofolini; Aldo Toni
Journal:  Int Orthop       Date:  2008-08-20       Impact factor: 3.075

2.  Experimental Validation of an ITAP Numerical Model and the Effect of Implant Stem Stiffness on Bone Strain Energy.

Authors:  K Ahmed; R J Greene; W Aston; T Briggs; C Pendegrass; M Moazen; G Blunn
Journal:  Ann Biomed Eng       Date:  2020-01-23       Impact factor: 3.934

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.