Noninvasive glucose measurement by monitoring of scattering coefficient during oral glucose tolerance tests. Non-Invasive Task Force.

L Heinemann1, U Krämer, H M Klötzer, M Hein, D Volz, M Hermann, T Heise, K Rave.   

Abstract

BACKGROUND: Continuous glucose monitoring by means of optical glucose sensors would allow patients with diabetes to check their metabolic control to their convenience. In an earlier study, we showed that noninvasive glucose monitoring is feasible for rapid changes in blood glucose by means of measuring the scattering coefficient of human skin. In this study, we investigated whether also slower changes in blood glucose, this time induced by an oral glucose load, can also be monitored by this approach.
METHODS: Five healthy subjects and 13 patients with type 2 diabetes have been given a 75-g oral glucose load. Portable noninvasive systems were used to measure the skin tissue scattering coefficient. For this purpose, two optical sensor heads were attached directly to the skin of each volunteer. Light was applied to the skin and the reflected light intensity was registered.
RESULTS: In 8 of 10 measurements, correlation of changes in scattering coefficient with changes in glycemia was acceptable. In 19 of 26 measurements (73%) of patients with type 2 diabetes the observed changes in the scattering coefficient also correlated in acceptable manner. The accordance between the simultaneous measurements of the two sensor heads was acceptable in 13 of 18 volunteers and patients studied. There were virtually no differences in the quality of the measurements between healthy volunteers and patients with diabetes.
CONCLUSIONS: This study shows that also slow changes in blood glucose induced by an oral glucose load can be monitored by registration of scattering coefficient changes. It remains to be elucidated why this has not been possible in all experiments.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 11469261     DOI: 10.1089/15209150050025168

Source DB:  PubMed          Journal:  Diabetes Technol Ther        ISSN: 1520-9156            Impact factor:   6.118


  3 in total

1.  Continuous noninvasive monitoring of changes in human skin optical properties during oral intake of different sugars with optical coherence tomography.

Authors:  Yuqing Zhang; Guoyong Wu; Huajiang Wei; Zhouyi Guo; Hongqin Yang; Yonghong He; Shusen Xie; Ying Liu
Journal:  Biomed Opt Express       Date:  2014-02-28       Impact factor: 3.732

2.  Improved CEEMDAN and PSO-SVR Modeling for Near-Infrared Noninvasive Glucose Detection.

Authors:  Xiaoli Li; Chengwei Li
Journal:  Comput Math Methods Med       Date:  2016-08-22       Impact factor: 2.238

Review 3.  Glucose biosensors in clinical practice: principles, limits and perspectives of currently used devices.

Authors:  Salvatore Andrea Pullano; Marta Greco; Maria Giovanna Bianco; Daniela Foti; Antonio Brunetti; Antonino S Fiorillo
Journal:  Theranostics       Date:  2022-01-01       Impact factor: 11.600

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.