Literature DB >> 11468196

Novel drug-delivery stent: intravascular ultrasound observations from the first human experience with the QP2-eluting polymer stent system.

Y Honda1, E Grube, L M de La Fuente, P G Yock, S H Stertzer, P J Fitzgerald.   

Abstract

BACKGROUND: The aim of this study was to use serial intravascular ultrasound (IVUS) to evaluate the long-term effect of stent-based 7-hexanoyltaxol (QP2, a taxane analogue) delivery on neointimal tissue growth within the stent and on vessel dimensions at the adjacent reference segments. METHODS AND
RESULTS: Serial IVUS analyses (immediately after intervention and at follow-up at 8.3 months) were performed in 15 native coronary lesions treated with the QuaDS-QP2 stent. IVUS measurements were performed at 8 cross-sections in each target segment (4 cross-sections within the stent and 2 cross-sections in each reference segment). At baseline, no significant plaque protrusion or thrombus was detected in the target segment. Mild incomplete stent apposition and edge dissection were observed in one and two cases, respectively. Percent expansion of the stent (minimum stent area/average reference lumen area) was 96.0+/-21.7%. At follow-up, mean neointimal area within the stent was 1.2+/-1.3 mm(2), and mean cross-sectional narrowing (neointimal area/stent area) was 13.6+/-14.9%. At the vessel segments immediately adjacent to the stent, a significant increase in plaque area (1.9+/-2.6 mm(2), P=0.001) was observed, but vessel area remained unchanged. However, no patients showed clinically significant in-stent or edge restenosis (diameter stenosis >/=50%) during the follow-up period.
CONCLUSIONS: The first human experience with the new drug-delivery stent showed a minimal amount of neointimal proliferation in the stented segment. Late lumen loss at the reference sites adjacent to the stent was acceptable and predominantly due to plaque proliferation.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11468196     DOI: 10.1161/hc2901.094149

Source DB:  PubMed          Journal:  Circulation        ISSN: 0009-7322            Impact factor:   29.690


  8 in total

Review 1.  Invasive assessment of the coronary circulation: intravascular ultrasound and Doppler.

Authors:  David E Newby; Keith A A Fox
Journal:  Br J Clin Pharmacol       Date:  2002-06       Impact factor: 4.335

2.  Evaluation of intracoronary stenting by intravascular optical coherence tomography.

Authors:  B E Bouma; G J Tearney; H Yabushita; M Shishkov; C R Kauffman; D DeJoseph Gauthier; B D MacNeill; S L Houser; H T Aretz; E F Halpern; I-K Jang
Journal:  Heart       Date:  2003-03       Impact factor: 5.994

Review 3.  Drug eluting stents: an updated meta-analysis of randomised controlled trials.

Authors:  C Roiron; P Sanchez; A Bouzamondo; P Lechat; G Montalescot
Journal:  Heart       Date:  2005-10-10       Impact factor: 5.994

Review 4.  Biomedical Imaging in Implantable Drug Delivery Systems.

Authors:  Haoyan Zhou; Christopher Hernandez; Monika Goss; Anna Gawlik; Agata A Exner
Journal:  Curr Drug Targets       Date:  2015       Impact factor: 3.465

5.  Usual dose of simvastatin does not inhibit plaque progression and lumen loss at the peri-stent reference segments after bare-metal stent implantation: a serial intravascular ultrasound analysis.

Authors:  Young Joon Hong; Myung Ho Jeong; Yun Ha Choi; Eun Hye Ma; Jum Suk Ko; Min Goo Lee; Keun Ho Park; Doo Sun Sim; Nam Sik Yoon; Hyun Ju Youn; Kye Hun Kim; Hyung Wook Park; Ju Han Kim; Youngkeun Ahn; Jeong Gwan Cho; Jong Chun Park; Jung Chaee Kang
Journal:  Korean J Intern Med       Date:  2010-11-27       Impact factor: 3.165

Review 6.  Drug-eluting stents: new era and new concerns.

Authors:  V Bhatia; R Bhatia; M Dhindsa
Journal:  Postgrad Med J       Date:  2004-01       Impact factor: 2.401

7.  Cardiotoxicities of paclitaxel in African Americans.

Authors:  Padma Kamineni; Kalpana Prakasa; Syed P Hasan; Ravi Akula; Fitzroy Dawkins
Journal:  J Natl Med Assoc       Date:  2003-10       Impact factor: 1.798

8.  Comparison of sirolimus and paclitaxel-eluting stents for complex coronary lesions: an intravascular ultrasound study.

Authors:  Yun-Kyeong Cho; Seung-Ho Hur; Hyun-Tae Kim; In-Cheol Kim; Hyoung-Seob Park; Hyuck-Jun Yoon; Chang-Wook Nam; Hyungseop Kim; Seong-Wook Han; Yoon-Nyun Kim; Kwon-Bae Kim
Journal:  Korean J Intern Med       Date:  2009-11-27       Impact factor: 2.884

  8 in total