Literature DB >> 11467555

Field-amplified sample injection combined with water removal by electroosmotic flow pump in acidic buffer for analysis of phenoxy acid herbicides by capillary electrophoresis.

L Zhu1, H K Lee.   

Abstract

A procedure that combines two common stacking techniques, field-amplified sample injection and water removal, with an electroosmotic flow pump, is used to separate phenoxy acid herbicides by capillary zone electrophoresis. Before sample loading, a long plug of water was hydrodynamically injected into the capillary both to serve as the medium to permit a high electric field strength and to contain sample anions. Because of this long length of water, the number of ions injected into the capillary was greatly increased. Electrokinetic injection at reversed voltage was then used for introducing negatively charged ions from the diluted sample into the column. The water was removed from the capillary using the electroosmotic flow (EOF) pump when the EOF of the background electrolyte was suppressed. This method afforded a sensitivity enhancement of greater than 3,000 times. Combined with solid-phase extraction, detection limits for the phenoxy acid herbicides as low as 0.01 ng/mL could be achieved.

Entities:  

Year:  2001        PMID: 11467555     DOI: 10.1021/ac001313f

Source DB:  PubMed          Journal:  Anal Chem        ISSN: 0003-2700            Impact factor:   6.986


  3 in total

1.  On-line sample preconcentration using field-amplified stacking injection in microchip capillary electrophoresis.

Authors:  Maojun Gong; Kenneth R Wehmeyer; Patrick A Limbach; Francisco Arias; William R Heineman
Journal:  Anal Chem       Date:  2006-06-01       Impact factor: 6.986

2.  Glycoform analysis of alpha1-acid glycoprotein based on capillary electrophoresis and electrophoretic injection.

Authors:  Chenhua Zhang; Cong Bi; William Clarke; David S Hage
Journal:  J Chromatogr A       Date:  2017-08-15       Impact factor: 4.759

3.  Combination of on-chip field amplification and bovine serum albumin sweeping for ultrasensitive detection of green fluorescent protein.

Authors:  Qiong Pan; Meiping Zhao; Shaorong Liu
Journal:  Anal Chem       Date:  2009-07-01       Impact factor: 6.986

  3 in total