Sources of uncertainty in isotope ratio measurements by inductively coupled plasma mass spectrometry.

P K Appelblad1, I Rodushkin, D C Baxter.   

Abstract

A model is presented describing the effects of dead time and mass bias correction factor uncertainties, flicker noise, and counting statistics on isotope ratio measurement precision using inductively coupled plasma mass spectrometry (ICPMS) with a single collector. Noise spectral analysis is exploited to enable estimation of the flicker noise parameters. For the instrument used, the flicker noise component exhibited a fairly weak frequency (t) dependence (is proportional to f -0.33+/-0.12), but was directly proportional to the total number of counts, Q. As white noise, determined by counting statistics, is given by Q0.5, the isotope ratio measurement uncertainties will actually cease to improve when Q exceeds a certain threshold. This would suggest that flicker noise could become the limiting factor for the precision with which isotope ratios can be determined by ICPMS. However, under most experimental conditions, uncertainties associated with mass discrimination and dead time correction factors are decisive. For ratios up to approximately 22 (115In/113In), optimum major isotope count rates are generally below 0.3 MHz, for which precision in the mass discrimination factor is limiting. The model derived could be used as a starting point for determining optimum conditions and understanding the limitations of single-collector ICPMS for precise isotope ratio measurements.

Year:  2001        PMID: 11467535     DOI: 10.1021/ac001537y

Source DB:  PubMed          Journal:  Anal Chem        ISSN: 0003-2700            Impact factor:   6.986


  3 in total

1.  Evaluation strategies and uncertainty calculation of isotope amount ratios measured by MC ICP-MS on the example of Sr.

Authors:  Monika Horsky; Johanna Irrgeher; Thomas Prohaska
Journal:  Anal Bioanal Chem       Date:  2016-01       Impact factor: 4.142

2.  The performance of single and multi-collector ICP-MS instruments for fast and reliable 34S/32S isotope ratio measurements.

Authors:  Ondrej Hanousek; Marion Brunner; Daniel Pröfrock; Johanna Irrgeher; Thomas Prohaska
Journal:  Anal Methods       Date:  2016-09-23       Impact factor: 2.896

3.  Comparison of Minimally Invasive Inductively Coupled Plasma-Mass Spectrometry Approaches for Strontium Isotopic Analysis of Medieval Stained Glass with Elevated Rubidium and Rare-Earth Element Concentrations.

Authors:  Alicia Van Ham-Meert; Eduardo Bolea-Fernandez; Joke Belza; Dan Bevan; Klaus Peter Jochum; Brigitte Neuray; Brigitte Stoll; Frank Vanhaecke; Line Van Wersch
Journal:  ACS Omega       Date:  2021-07-09
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.