| Literature DB >> 11466788 |
T Donnelly1, S Waldron, A Tait, J Dougans, S Bearhop.
Abstract
This paper describes the application of a simple chromium reduction furnace which can be interfaced with a dual inlet isotope-ratio mass spectrometer thus providing the capacity for cheap, fast, accurate and precise measurement of deltaD(V-SMOW) by dynamic mass spectrometry. Measurements are precise to the order of < or =0.5 per thousand. Mean 95% confidence intervals for the Vienna Standard Mean Ocean Water (V-SMOW) to Standard Light Antarctic Precipitation (SLAP) range are in the order of 2.5 per thousand and the system is linear over the range -428 to 23,000 per thousand. Memory effects do exist, but are small for natural abundance samples and can be minimised by careful planning of the analytical load. Copyright 2001 John Wiley & Sons, Ltd.Entities:
Mesh:
Substances:
Year: 2001 PMID: 11466788 DOI: 10.1002/rcm.361
Source DB: PubMed Journal: Rapid Commun Mass Spectrom ISSN: 0951-4198 Impact factor: 2.419