| Literature DB >> 11461010 |
M Gilar1, A Belenky, B H Wang.
Abstract
In the last 10 years mass spectrometry (MS) has become an important method for analysis of peptides, proteins and DNA. It was recently utilized for accurate high-throughput protein identification, sequencing and DNA genotyping. The presence of non-volatile buffers compromises sensitivity and accuracy of MS biopolymer analysis; it is essential to remove sample contaminants prior to analysis. We have developed a fast and efficient method for desalting of DNA oligonucleotides and peptides using 96-well solid-phase extraction plates packed with 5 mg of Waters Oasis HLB sorbent (Waters, Milford, MA, USA). This reversed-phase sorbent retains the biopolymer analytes, while non-retained inorganic ions are washed out with pure deionized water. DNA oligonucleotides or peptides are eluted using a small amount (20-100 microl) of acetonitrile-water (70:30, v/v) solution. The SPE desalting performance meets the requirements for MS applications such as protein digest analysis and DNA genotyping.Entities:
Mesh:
Substances:
Year: 2001 PMID: 11461010 DOI: 10.1016/s0021-9673(01)00833-0
Source DB: PubMed Journal: J Chromatogr A ISSN: 0021-9673 Impact factor: 4.759