| Literature DB >> 1145902 |
K E Arfors, D Bergqvist, M Intaglietta, B Westergren.
Abstract
A method for the measurement of red blood cell velocity in the microcirculation is described. The technique is based on the on-line cross-correlation of the delay between optical signatures derived from photodetectors aligned in direction oflow. A specialised photodiode holder which can be inserted directly in place of an ocular is described. The interdetector delay data is computed automatically from the cross-correlation function by a peak detector designed to provide a voltage proportional to the delay to maximum cross-correlation, regardless of intermediate lower peaks. Application of this system to measurements in the rabbit ear chamber, mesentery and skeletal muscle is described.Entities:
Mesh:
Year: 1975 PMID: 1145902 DOI: 10.3109/03009737509178987
Source DB: PubMed Journal: Ups J Med Sci ISSN: 0300-9734 Impact factor: 2.384