Comparison of capillary and venous glucose measurements in healthy volunteers.

D L Funk1, L Chan, N Lutz, V P Verdile.   

Abstract

BACKGROUND: Out-of-hospital (OOH) emergency personnel measure serum glucose in order to determine the need for dextrose therapy. Most devices that measure serum glucose are designed to use capillary blood obtained from a finger puncture. However, OOH emergency personnel often use venous blood obtained during intravenous line (IV) placement to determine serum glucose.
OBJECTIVE: To compare capillary and venous glucose measurements.
METHODS: This prospective study used healthy, non-fasting volunteers. Simultaneous venous and capillary blood samples were obtained from each subject. Glucose levels were measured using a glucometer designed for capillary samples. The capillary and venous measurements were compared using a Pearson correlation coefficient. Power analysis revealed that the study had the ability to detect a difference of 15 mg/dL.
RESULTS: Ninety-seven volunteers (56 males, 41 females) with a mean age of 37 +/- 11.9 years were enrolled. The mean capillary and venous glucose values were 104.5 +/- 20.7 mg/dL and 109.7 +/- 22.4 mg/dL, respectively. The Pearson correlation coefficient was 0.24.
CONCLUSIONS: The correlation between venous and capillary blood glucose measurements is relatively poor in this group of healthy volunteers. Further research must be conducted on patients at risk for abnormal blood glucose.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11446542     DOI: 10.1080/10903120190939788

Source DB:  PubMed          Journal:  Prehosp Emerg Care        ISSN: 1090-3127            Impact factor:   3.077


  4 in total

1.  Capillary versus venous bedside blood glucose estimations.

Authors:  R Boyd; B Leigh; P Stuart
Journal:  Emerg Med J       Date:  2005-03       Impact factor: 2.740

2.  Glycemic control and the outcomes of Hispanic patients with diabetes admitted to the general ward of a community hospital in Puerto Rico.

Authors:  Nancy Torres-Torres; Miguel A Maldonado-Rodríguez; Shirley Pérez-López; Kassandra Sierra-Martínez; Astrid J García
Journal:  P R Health Sci J       Date:  2011-06       Impact factor: 0.705

3.  Comparison of Point of Care (POC) Testing of Glucose by B Braun Glucometer and Hemocue Glucose 201+ Analyser Versus Centralised Testing in Neonatal Intensive Care Unit (NICU).

Authors:  Sudha Reddy V R; Sumathi M E; Beere Gowda Y C; Mohamed Suhail S
Journal:  J Clin Diagn Res       Date:  2014-07-20

4.  Comparison of capillary and venous blood glucose levels using glucometer and laboratory blood glucose level in poisoned patients being in coma.

Authors:  Ahmad Yaraghi; Nastaran Eizadi Mood; Leila Kamali Dolatabadi
Journal:  Adv Biomed Res       Date:  2015-11-23
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.