Axial characteristics of circular external skeletal fixator single ring constructs.

D D Lewis1, D G Bronson, A R Cross, R D Welch, P S Kubilis.   

Abstract

OBJECTIVE: Evaluate the effects of varying ring diameter, wire tension, and wire-divergence angle on the axial stiffness characteristics of circular external skeletal fixator single-ring constructs. Study Design-Biomechanical evaluation using circular fixator components and a Delrin cylinder bone model.
METHODS: Single ring constructs using two 1.6 mm diameter Kirschner wires to secure a 19 mm Delrin cylinder centered within the ring were examined. Component variables evaluated were ring diameter (50 mm, 66 mm, 84 mm, and 118 mm), wire-divergence angle (30 degrees, 60 degrees, and 90 degrees ), and wire tension (0 kg, 30 kg, 60 kg, and 90 kg). A total of 48 constructs were examined. Rings were rigidly mounted on a universal testing system and the cylinder loaded in axial compression (7.4 N/s) to 220 N. Load/displacement curves were analyzed to determine the following: the displacement (mm) that occurred before the slope of each load/displacement curve became linear, the stiffness (N/mm) of the linear portion of each load/deformation curve, and the total displacement (mm) produced at maximal load. Least-squares linear regression was used to model response variables as linear functions of ring diameter, wire divergence angle, and wire tension. Three-way interactions and 2-way interactions among independent component variables were evaluated first in the modeling process and included in a best model if response variables were found to have statistically significant regression coefficients. The regression coefficients and corresponding standard errors and covariances were used to estimate the maximal effect and standard error attributable to wire divergency angle (change from 30 degrees to 90 degrees ) and wire tension (change from 0 to 90 kg) for each ring diameter.
RESULTS: All load/deformation curves had an initial exponential increase in stiffness, with the slope becoming linear at higher loads. The exponential phase was more pronounced in larger-diameter ring constructs and was mitigated by tensioning the wires. Ring diameter had the greatest influence on displacement that occurred before the curve became linear (semipartial r(2) [sp-r2] = .89), stiffness (sp-r2 = .94), and total displacement (sp-r2 = .93). Wire tension exerted a smaller influence on displacement that occurred before the curve became linear (sp-r2 =.06), stiffness (sp-r2 = .03), and total displacement (sp-r2 = .05). Wire divergence angle had a nominal effect on displacement that occurred before the curve became linear (sp-r2 = .0001), on stiffness (sp-r2 = .004), and on total displacement (sp-r2 =.003).
CONCLUSIONS: Ring diameter had a profound effect on the axial stiffness characteristic of single ring constructs. Tensioning of the fixation wires can improve the axial stiffness characteristics of these constructs, particularly in larger diameter ring constructs, by mitigating the initial exponential phase of the load/deformation curve. Wire divergence angle had only a nominal differential effect on axial stability. CLINICAL RELEVANCE: Understanding how individual component variables and their interactions influence bone segment stability should help surgeons to optimize interfragmentary strain. Tensioning fixation wires is probably unnecessary in 50 mm diameter ring constructs, but assumes greater importance as ring diameter increases. Copyright 2001 by The American College of Veterinary Surgeons

Mesh:

Year:  2001        PMID: 11443601     DOI: 10.1053/jvet.2001.24397

Source DB:  PubMed          Journal:  Vet Surg        ISSN: 0161-3499            Impact factor:   1.495


  6 in total

1.  What Are the Biomechanical Properties of the Taylor Spatial Frame™?

Authors:  Daniel J Henderson; Jeremy L Rushbrook; Paul J Harwood; Todd D Stewart
Journal:  Clin Orthop Relat Res       Date:  2016-11-28       Impact factor: 4.176

2.  What Are the Biomechanical Effects of Half-pin and Fine-wire Configurations on Fracture Site Movement in Circular Frames?

Authors:  Daniel J Henderson; Jeremy L Rushbrook; Todd D Stewart; Paul J Harwood
Journal:  Clin Orthop Relat Res       Date:  2015-12-07       Impact factor: 4.176

3.  Descriptive report of antebrachial growth deformity correction in 17 dogs from 1999 to 2007, using hybrid linear-circular external fixator constructs.

Authors:  Colin W Sereda; Daniel D Lewis; Robert M Radasch; Charles W Bruce; Kristin A Kirkby
Journal:  Can Vet J       Date:  2009-07       Impact factor: 1.008

4.  The use of circular and hybrid external skeletal fixation systems to repair open tibial fractures in large ruminants: a report of six clinical cases.

Authors:  Mudasir Ahmad Shah; Rohit Kumar; Prakash Kinjavdekar; H P Aithal; Mohammad Arif Basha; Asif Majid
Journal:  Vet Res Commun       Date:  2022-02-04       Impact factor: 2.459

5.  Comparison of transverse wires and half pins in Taylor Spatial Frame: a biomechanical study.

Authors:  Ashish Khurana; Carlton Byrne; Sam Evans; Hiro Tanaka; Kartik Haraharan
Journal:  J Orthop Surg Res       Date:  2010-03-27       Impact factor: 2.359

6.  Evaluation of sixty-eight cases of fracture stabilisation by external hybrid fixation and a proposal for hybrid construct classification.

Authors:  María Jiménez-Heras; Gian Luca Rovesti; Gianluca Nocco; Massimo Barilli; Paolo Bogoni; Emilio Salas-Herreros; Matteo Armato; Francesco Collivignarelli; Federico Vegni; Jesus Rodríguez-Quiros
Journal:  BMC Vet Res       Date:  2014-09-20       Impact factor: 2.741

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.