Literature DB >> 11442463

Detection of myocardial ischemia by electron beam CT. Experimental studies.

F D Knollmann1, P Muschick, W Krause, H Hausmann, R Hetzer, R Felix.   

Abstract

PURPOSE: To determine if contrast-enhanced electron beam CT (EBCT) can detect areas of acute myocardial ischemia, and if pharmacological stress testing improves the diagnostic accuracy of EBCT.
MATERIAL AND METHODS: We injected 0.5 ml/kg and 1.0 ml/kg b.w. of iopromide at a rate of 4 ml/s into the right atrium of 5 ventilated female minipigs at rest and after occlusion of the left anterior descending (LAD) coronary artery. Both ventricles were examined at six short axis levels with an EBCT unit. Myocardial perfusion was calculated from the time-density curves of four left ventricular myocardial segments and the aorta. We also tested the effect of the contrast agent on myocardial density after i.v. administration of 0.6 mg/kg dipyridamole before and after LAD occlusion.
RESULTS: At rest, the contrast agent increased myocardial density by 28+/-2 HU, corresponding to a myocardial perfusion estimate of 67+/-7 ml/min/100 g. After dipyridamole, myocardial density increased by 29+/-4 HU. Following occlusion of the LAD, anteroseptal myocardium displayed 10+/-4 HU density increase. The area of non-enhancement corresponded to ischemic myocardium in stained pathologic sections.
CONCLUSION: Contrast-enhanced stress EBCT can be used to detect areas of myocardial ischemia, and EBCT stress perfusion imaging may be necessary to consistently differentiate ischemic from non-ischemic myocardial tissue.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11442463     DOI: 10.1080/028418501127347034

Source DB:  PubMed          Journal:  Acta Radiol        ISSN: 0284-1851            Impact factor:   1.701


  3 in total

1.  CT coronary angiography: quantitative assessment of myocardial perfusion using test bolus data-initial experience.

Authors:  Ashley M Groves; Vicky Goh; Sabarinath Rajasekharan; Irfan Kayani; Raymondo Endozo; John C Dickson; Leon J Menezes; Manu Shastry; Said B Habib; Peter J Ell; Brian F Hutton
Journal:  Eur Radiol       Date:  2008-05-09       Impact factor: 5.315

2.  Detection of myocardial perfusion abnormalities: standard dual-source coronary computed tomography angiography versus rest/stress technetium-99m single-photo emission CT.

Authors:  W Cheng; M Zeng; C Arellano; W Mafori; J Goldin; M Krishnam; S G Ruehm
Journal:  Br J Radiol       Date:  2010-04-22       Impact factor: 3.039

3.  Early and delayed myocardial enhancement in myocardial infarction using two-phase contrast-enhanced multidetector-row CT.

Authors:  Sung-Min Ko; Young-Whan Kim; Seong-Wook Han; Joon-Beom Seo
Journal:  Korean J Radiol       Date:  2007 Mar-Apr       Impact factor: 3.500

  3 in total