| Literature DB >> 11434533 |
M Rich1, R M Bannatyne, P Topham, Z Memish.
Abstract
Using a straight-from-the-bottle into-the-instrument technique, a series of direct susceptibility tests were performed with the BACTEC 9240 and Microscan Walkaway systems. Although rapid, unacceptably high error rates were obtained when compared with a standardized method. This procedure is not sufficiently accurate for clinical use.Mesh:
Year: 2001 PMID: 11434533 DOI: 10.1080/1120009x.2001.11782332
Source DB: PubMed Journal: J Chemother ISSN: 1120-009X Impact factor: 1.714